R&S ZVL
Command Reference
Instrument-Control Commands
Operating Manual 1303.6580.32-05
451
Example:
H
COP:PAGE:ORI LAND; HCOP
Select landscape page orientation and start printing.
HCOPy:PAGE:WINDow ACTive | ALL | SINGle
Defines the number of diagram areas per printed page.
ACTive
Print only active diagram area.
ALL
All diagram areas are printed on one page.
SINGle
One diagram area per page.
*RST value
ALL
SCPI, Command Types
Confirmed, command or query
Example:
H
COP:PAGE:WIND SING; HCOP
Select one diagram area per page and start printing.
INITiate<Ch>...
This subsystem controls the initiation of the trigger system and defines the scope of the triggered
measurement.
INITiate<Ch>:CONTinuous <Boolean>
Qualifies whether the analyzer measures in single sweep or in continuous sweep mode.
<Ch>
Channel number. The channel is ignored because the command affects all
channels.
<Boolean>
ON - The analyzer measures continuously, repeating the current sweep.
OFF - The measurement is stopped after the number of sweeps defined via
[SENSe<Ch>:]SWEep:COUNt
. INITiate<Ch>[:IMMediate] initiates a new
measurement cycle.
*RST value
ON
SCPI,
Command
Types
Confirmed, command or query
Example:
*RST; INIT:CONT OFF
Activate single sweep mode for all channels (including channel no. 2 created
later).
INIT:SCOP SING
State that a single sweep will be performed in the active channel only.
CALC2:PAR:SDEF 'TRC2','S11'; INIT2