The S-parameters are the basic measured quantities of a network analyzer. They describe how the DUT
modifies a signal that is transmitted or reflected in forward or reverse direction. S-parameters are
expressed as S
<out>< in>
, where <out> and <in> denote the output and input port numbers of the DUT.
A full 2-port S-parameter measurement involves 2 stages (referred to as partial measurements) with
interchanged drive and receive ports. The analyzer automatically switches the internal power sources and
the receivers to obtain the desired S-parameters
.
Remote
control:
CALCulate<Ch>:PARameter:MEASure "<Trace_Name>", "S11" | "S12" |
"S21" | "S22"
[SENSe<Chn>:]FUNCtion[:ON] "...:POWer:S<11 | 12 | 21 | 22>"
Create new trace and select name and measurement parameter:
CALCulate<Ch>:PARameter:SDEFine "<Trace_Name>", "S11" | "S12" |
"S21" | "S22"
S-Parameter Wizard
The
S-Parameter Wizard
consists of a series of dialogs providing the settings for a standard S-parameter
measurement.
The measurement comprises the following stages:
Select the measurement parameters and the diagram areas
Choose an S-parameter or a group of S-parameters to be measured and displayed. The
measurement parameters are displayed in separate diagram areas: Cartesian diagram areas are