R&S ZVL
Command Reference
Instrument-Control Commands
Operating Manual 1303.6580.32-05
510
<port_no>
Port number of the analyzer. This numeric suffix is ignored; the command
affects the parameters of all ports.
<Boolean>
The parameter function depends on whether the command is used as a setting
command or as a query:
For setting command:
ON – no effect
OFF – resets all length offsets to zero
For query:
1 – at least one length offset is different from zero
0 – all length offsets are zero
*RST value
OFF
SCPI, Command
Types
Device-specific, command or query
Example:
*RST; CORR:OFFS?
Reset the instrument and query whether the length offset parameters have
been reset as well. The response is
0
.
[SENSe<Ch>:]CORRection:OFFSet<port_no>:MAGNitude <DC_loss>
Defines the frequency-independent part (DC value) of the offset loss.
Use the
[SENSe<Ch>:]CORRection:LOSS<port_no>...
commands to define the complete set of
loss offset parameters.
[SENSe<Ch>:]CORRection:OFFSet<port_no>:MAGNitude
is equivalent to
[SENSe<Ch>:]CORRection:LOSS<port_no>:OFFSet
<Ch>
Channel number of the offset-corrected channel
<port_no>
Port number of the analyzer
<DC_loss>
Frequency-independent part of the offset loss
Range [def. unit]
-200 dB to +200 dB [dB]. The increment (UP/DOWN) is 0.001 dB.
*RST value
0 dB
SCPI, Command Types
Confirmed, command or query
[SENSe<Ch>:]CORRection:SSTate?
Returns the state of the active system error correction for channel <Ch>.
<Ch>
Channel number of the calibrated channel.
Response
Sting variable containing the calibration state label in the trace list ('Cal', 'Cai',
'Ca?', 'Cav', 'Cal Off').
*RST value
'' (empty string)
SCPI, Command
Types
Device-specific, query only
Example:
See [SENSe<Ch>:]CORRection:DATE?