R&S ZVL
Command Reference
Instrument-Control Commands
Operating Manual 1303.6580.32-05
528
Example:
SEGM:ADD
Create a new sweep segment no. 1 in channel no. 1 using default settings and thus
0 s meas. delay.
SEGM:SWE:DWELl 0.1
Increase the meas. delay to 0.1 s.
SEGM:SWE:DWELl:CONT OFF
Couple the meas. delay in all segments and reset the delay in segment no. 1 to the
initial value of 0 s.
[SENSe<Ch>:]SEGMent<Seg>:SWEep:POINts <Points>
Defines the total number of measurement
Points
in sweep segment no. <Seg>.
<Ch>
Channel number.
<Seg>
Sweep segment number.
<Points>
Number of points in the segment.
Range [def. unit]
1 to 2147483647. [1] 1 is allowed if start and stop frequencies are equal.
*RST value
51
SCPI, Command
Types
Device-specific, command or query
Example:
SEGM:ADD
Create a new sweep segment no. 1 in channel no. 1 using default settings
and thus 51 sweep points.
SEGM:SWE:POIN 401
Increase the number of points to 401.
[SENSe<Ch>:]SWEep...
This subsystem provides general settings to control the sweep . Most of the settings are related to the
sweep time.
The [SENSe<Ch>:]SWEep... settings apply to non-segmented sweep types. Segmented sweeps are
selected via
[SENSe<Ch>:]SWEep:TPYE SEGMent
and configured in the
[SENSe<Ch>:]SEGMent<Seg>... subsystem.
[SENSe<Ch>:]SWEep:COUNt <No_of_Sweeps>
Defines the number of sweeps
to be measured in single sweep mode (
INITiate<Ch>:CONTinuous
OFF
).
<Ch>
Channel number.
<No_of_Sweeps>
Number of consecutive sweeps to be measured.