[SENSe<Ch>:]CORRection[:STATe] <Boolean>
Enables or disables the system error correction for channel <Ch>.
<Ch>
Calibrated channel number
<Boolean>
Enables (ON) or disables (OFF) the correction.
*RST value
ON [–]
SCPI, Command
Types
Confirmed, command or query
Example:
*RST; CORR?
Reset the instrument and query whether channel 1 is system error corrected.
The response is
1
.
[SENSe<Ch>:]FREQuency...
This subsystem sets frequency-related parameters, especially the measurement and display ranges for
the different sweep types.
The frequency ranges for the different instrument models are listed below. The specified frequency ranges
start at 9 kHz; see data sheet.
R&S ZVL3 and R&S ZVL3-75 R&S ZVL6
R&S ZVL13
Start, Stop
10 Hz to 3 GHz
10 Hz to 6 GHz
5 kHz to 15 GHz
The stop freq. is preset to 13.6 GHz;
see data sheet
Center
> 10 Hz to < 3 GHz
> 10 Hz to < 6 GHz
> 5 kHz to < 15 GHz
Span
10 Hz to 2.99999999 GHz
10 Hz to 5.99999999 GHz 10 Hz to 14.99999999 GHz
Note:
The sweep range can be defined alternatively by a combination of Start/Stop frequencies or a
Center frequency and Span.
[SENSe<Ch>:]FREQuency:CENTer <center_frequency>
Defines the center of the measurement and display range for a frequency sweep.
<Ch>
Channel number. If unspecified the numeric suffix is set to 1.
<center_frequency>
Center frequency of the sweep.
Range [def. unit]
Depending on the instrument model [Hz]. The increment (parameters UP or
DOWN) is 0.1 kHz.
*RST value
Center of the analyzer's default frequency range: (f
start
+ f
stop
)/2
SCPI, Command
Types
Confirmed, command or query.
Example:
FUNC "XFR:POW:RAT B1, A2"
Activate a frequency sweep and select the ratio B1/A2 as measured
parameter for channel and trace no. 1.
FREQ:CENT 100MHz