R&S ZVL
GUI Reference
Channel Menu
Operating Manual 1303.6580.32-06
262
Trigger Settings
Trigger Settings
opens a dialog to specify the operation that the analyzer carries out after receiving a
trigger event.
The dialog is divided into the following panels (for a detailed description refer to the following sections):
•
Triggered Measurement Sequence
selects the measurement cycle to be triggered.
•
Trigger Delay
specifies a delay time between the trigger event and the start of the measurement.
The Trigger Settings are not valid for Free Run measurements.
Triggered Measurement Sequence
Selects the measurement cycle to be triggered.
•
Sweep
means that each trigger event starts an entire sweep, according to the current sweep
configuration.
•
Point
means that each trigger event starts the measurement at the next sweep point. This mode is
not available on R&S ZVL13 network analyzers.
The period of the trigger events should be adjusted to the triggered measurement sequence. If the
analyzer receives a trigger event while the last sequence is still running, the trigger event is
skipped and a message Trigger event missed is displayed.
Remote control:
TRIGger[:SEQuence]:LINK
Trigger Delay
Specifies a delay time between the trigger event and the start of the measurement.
The delay time entered must be zero or positive so that the trigger event precedes the start of the
measurement (post-trigger).
Remote control:
TRIGger[:SEQuence]:HOLDoff
Average Factor
Opens the numeric entry bar to define the number of consecutive sweeps to be averaged.