R&S ZVL
Command Reference
Instrument-Control Commands
Operating Manual 1303.6580.32-05
359
Types
Example:
*RST; CALC:FILT:TIME:STAT ON
Reset the instrument and enable the time gate.
CALC:FILT:TIME NOTCh
Select a notch filter in order to reject unwanted pulses.
CALCulate<Chn>:FILTer[:GATE]:TIME:WINDow RECT | HAMMing | HANN |
BOHMan | DCHebyshev
Selects the time gate to be applied to the time domain transform.
<Chn>
Channel number used to identify the active trace
Parameters
RECT – Steepest edges (rectangle)
HANN – Normal gate (Hann)
HAMMing – Steep edges (Hamming)
BOHMan – Minimum flatness (Bohman)
DCHebyshev – Arbitrary gate shape (Dolph-Chebychev)
*RST value
HANN
SCPI, Command Types
Confirmed (with some device-specific parameters), command or query
Example:
See CALCulate<Chn>:FILTer[:GATE]:TIME:DCHebychev
CALCulate<Chn>:GDAPerture...
This subsystem defines parameters for the group delay calculation.
CALCulate<Chn>:GDAPerture:SCOunt <steps>
Defines the number of aperture steps for the group delay calculation.
<Chn>
Channel number used to identify the active trace. If unspecified the numeric
suffix is set to 1.
<steps>
Aperture steps
Range [def. unit]
1 to 10000
*RST value
10
SCPI, Command
Types
Device-specific (last keyword), with query
Example:
*RST; CALC:FORMat GDELay
Select group delay format for the default trace.
CALC:GDAP:SCO 12
Select 12 aperture steps for the calculation of the delay at each sweep point.