CALCulate<Chn>:LIMit:CONTrol:SHIFt <offset>
Shifts the limit line in horizontal direction.
<Chn>
Channel number used to identify the active trace.
<offset>
Amount that the limit line is shifted.
Range [def.
unit]
Almost no restriction for limit segments; see
Rules for Limit Line Definition.
[Hz.
For distance-to-fault traces the default unit is the length unit defined via
UNIT:LENGth
.]]
*RST value
–
SCPI,
Command
Types
Device-specific, no query.
Example:
*RST; CALC:LIM:CONT 1 GHZ, 2 GHZ
Define a limit line segment in the stimulus range between 1 GHz and 2 GHz, using
default response values.
CALC:LIM:CONT:SHIF 1; CALC:LIM:CONT?
Shift the segment by 1 Hz. The modified limit line segment extends from
1000000001 (Hz) to 2000000001 (Hz).
CALCulate<Chn>:LIMit:DATA <type>, <start stim>, <stop stim>, <start resp>,
<stop resp>{,<type>, <start stim>, <stop stim>, <start resp>, <stop resp>}
Defines the limit line type, the stimulus and response values for a limit line with an arbitrary number of limit
line segments.
In contrast to CALCulate<Chn>:LIMit:CONTrol[:DATA], this command does not overwrite existing limit
line segments. The defined segments are appended to the segment list as new segments.
<Chn>
Channel number used to identify the active trace.
<type>
Identifier for the type of the limit line segment.
Range [def.
unit]
0
– Limit line segment off, segment defined but no limit check performed
1
– Upper limit line segment
2
– Lower limit line segment
<start stim>,
<stop stim>,
<start resp>,
<stop resp>
Stimulus and response values of the first and last points of the limit line segment.
Range [def.
unit]
Almost no restriction for limit segments; see
Rules for Limit Line Definition.
[no unit
entries accepted, the default unit is Hz. For distance-to-fault traces the default unit
is the length unit defined via
UNIT:LENGth
.]
*RST values
– (no limit line defined after *RST)
SCPI,
Command
Types
Device-specific, with query.
Example:
*RST; CALC:LIM:CONT 1 GHZ, 1.5 GHZ
Define an upper limit line segment in the stimulus range between 1 GHz and 1.5