R&S ZVL
Command Reference
Instrument-Control Commands
Operating Manual 1303.6580.32-05
373
Use
CALCulate<Chn>:LIMit:STATe
to switch on or off the entire limit check, including upper and
lower limit lines.
<Chn>
Channel number used to identify the active trace.
<Boolean>
ON | OFF - Limit check on or off.
*RST value
OFF
SCPI, Command Types
Confirmed, command or query.
Example:
CALC:LIM:UPP -10, 0, 0, -10
Define the following upper and (default) lower limit line segments:
CALC:LIM:UPP:STAT ON; CALC:LIM:FAIL?
Switch the limit check on and query the result.
CALCulate<Chn>:MARKer<Mk>...
This subsystem controls the marker functions. The commands are device-specific and beyond what is
specified in the SCPI subsystem
SOURce:MARKer
.
CALCulate<Chn>:MARKer<Mk>:AOFF
Removes all markers from all traces of the active setup. The removed markers remember their properties
(stimulus value, format, delta mode, number) when they are restored (CALC<Chn>:MARK<Mk> ON). The
marker properties are definitely lost if the associated trace is deleted.
<Chn>
Channel number used to identify the active trace. If unspecified the numeric
suffix is set to 1.
<Mk>
Marker number in the range 1 to 10. This numeric suffix is ignored and may
be set to any value.
*RST value
–
SCPI, Command
Types
Device-specific, no query.
Example:
Suppose that the active setup contains an active trace no. 1.
CALC:MARK1 ON; MARK2 ON
Create markers 1 and 2 and assign them to the trace no. 1.
CALC:MARK:AOFF
Remove both markers.