R&S ZVL
Preparing for Use
Putting the Instrument into Operation
Quick Start Guide 1303.6538.62 - 06
23
Danger of injury
To avoid injuries, place the instrument on a stable surface and do not stack other
instruments or material on top of it.
1.3.4 Mounting in a 19" Rack
Using the adapter R&S ZZA-S334 (order no. 1109.4487.00) the instrument can be
mounted in 19" racks according to the mounting instructions supplied with the rack
adapter.
C
Allow for sufficient air supply in the rack.
C
Make sure that there is sufficient space between the ventilation holes and the rack
casing.
1.3.5 EMI Protective Measures
In order to avoid electromagnetic interference, the instrument may only be operated
when it is closed and with all shielding covers fitted. Only appropriate shielded signal
and control cables may be used.
1.3.6 Power Supply Options
The R&S ZVL is equipped with an AC power supply connector. In order to use the
analyzer independently from an AC power supply, it can be fitted with a DC power
supply connector (option DC Power Supply, R&S FSL–B30) and/or a battery pack
(option NIMH Battery Pack, R&S FSL–B31). Refer toDC Power Supply and Battery for
further information.
From the available power supplies, the R&S ZVL selects the one to use according to
the following priority scheme:
Priority
Power Supply
1
AC power supply
2
DC power supply
3
Battery