Remote Control
Remote Control Operation
Operating Manual 1303.6580.32-05
318
The analyzer provides a third display option where the measurement screen is only updated
when this is triggered by a remote control command; see
SYSTem:DISPlay:UPDate ONCE
.
The instrument remains in the remote state until it is reset to the manual state via the GUI or via GPIB bus
(see section
Return to Manual Operation)
.
You can create your own keys to replace the two default softkeys Go to Local and Display
On/Off. See Combining Manual and Remote Control.
A tooltip across the bottom of the remote screen indicates a remote command error, e.g.
. You can switch off this tooltip using
SYSTem:ERRor:DISPlay OFF
.
Remote control:
@REM
SYSTem:DISPlay:UPDate
SYSTem:USER:DISPlay:TITLe
(define a title for the remote
display)
SYSTem:ERRor:DISPlay
(switch tooltip on or off)
Setting the Device Address
The GPIB address (primary address) of the instrument is factory-set to 20. It can be changed manually in
the
System
menu or via GPIB bus. For remote control, addresses 0 through 30 are permissible. The GPIB
address is maintained after a reset of the instrument settings.
Remote control:
SYSTem:COMMunicate:GPIB[:SELF]:ADDRess
Return to Manual Operation
Return to manual operation can be initiated via the front panel or via remote control.
Manually: Click the
Local
softkey in the remote screen.
Via GPIB bus: CALL IBLOC(device%)
Via RSIB or VX 11 protocol:
@LOC
and
@REM
can be used to switch from remote to manual
control and back.
Local lockout
Before returning to manual control, command processing must be completed. If this is not the case, the
analyzer switches back to remote control immediately.
Returning to manual control by pressing the
Local
softkey can be disabled by the GPIB Local Lockout
Message (LLO; see
GPIB Bus Interface, Universal Commands)
which is also included in the NI
commands SetRWLS (Set Remote With Lockout State) or SendLLO. This prevents unintentional switch-
over, i.e. return to manual control is possible via the GPIB bus only.
Returning to manual control via the front panel keys can be enabled again by deactivating the REN control
line of the GPIB bus.