Create channel 4 and a trace named
Ch4Tr1
to measure the input reflection
coefficient S11. The trace becomes the active trace in channel 4.
CALC4:FORM MLIN; DISP:WIND:TRAC:FEED 'CH4TR1'
Calculate the magnitude of S11 and display it in a linearly scaled Cartesian
diagram.
Assume that the result at a sweep point is given by the complex quantity z = x + jy. The meaning of the
parameters is as follows (see also table in
CALCulate<Chn>:MARKer<Mk>:FORMat
description):
MLINear
Calculate the magnitude of z, to be displayed in a Cartesian diagram with a linear scale
MLOGarithmic
Calculate the magnitude of z, displayed in a Cartesian diagram with a logarithmic scale
PHASe
Phase of z, displayed in a Cartesian diagram with a linear vertical axis
UPHase
Unwrapped phase of z, displayed in a Cartesian diagram with a linear vertical axis
POLar
Magnitude and phase, displayed in a polar diagram
SMITh
Magnitude and phase, displayed in a Smith chart
ISMith
Magnitude and phase, displayed in an inverted Smith chart
GDELay
Group delay, displayed in a Cartesian diagram
REAL
Real part (x), displayed in a Cartesian diagram
IMAGinary
Imaginary part (y), displayed in a Cartesian diagram
SWR
Standing Wave Ratio (SWR), displayed in a Cartesian diagram
COMPlex
x, y, displayed in a polar diagram
MAGNitude
Magnitude (sqrt(x
2
+ y
2
)), displayed in a Cartesian diagram with a logarithmic scale
CALCulate<Chn>:FILTer[:GATE]...
This subsystem defines the properties of the time gate optimize the time domain response (with option
ZVL-K3,
Time Domain)
.
CALCulate<Chn>:FILTer[:GATE]:TIME:CENTer <numeric_value>
Defines the center time of the time gate.
<Chn>
Channel number used to identify the active trace
<numeric_value>
Center time of the time gate
Range [def. unit]
-99.8999999 s to +99.8999999 s. The increment (parameters UP or
DOWN) is 0.1 ns. [s]
*RST value
1.5E-009 s
SCPI, Command
Types
Confirmed, command or query
Example:
*RST; CALC:TRAN:TIME:STAT ON; CALC:FILT:TIME:STAT ON
Reset the instrument and enable the time domain representation and the
time gate.