R&S ZVL
Command Reference
Instrument-Control Commands
Operating Manual 1303.6580.32-05
376
Create a reference marker at the center of the sweep range and set marker 1
to delta mode.
CALCulate<Chn>:MARKer<Mk>:FORMat DEFault | MLINear | MLOGarithmic |
PHASe | POLar | COMPlex | GDELay | REAL | IMAGinary | SWR | LINPhase |
LOGPhase | IMPedance | ADMittance | MDB | MLPHase | MDPHase
Defines the output format for the (complex) value of marker <Mk> on trace no. <Chn>.
Note:
The formats of the markers assigned to a trace are independent of each other and of the trace
format settings; see
CALCulate<Chn>:FORMat
.
<Chn>
Channel number used to identify the active trace.
<Mk>
Marker number in the range 1 to 10.
Parameters
See list of parameters below.
*RST value
DEFault
SCPI, Command
Types
Device-specific, command or query
Example:
Suppose that the active setup contains an active trace no. 1.
CALC:MARK ON; CALC:MARK:FORM?
Create marker 1, assign it to the trace no. 1 and query its format. The analyzer
returns the format of the active trace.
Assume that the marker result is given by the complex quantity z = x + jy. The meaning of the parameters
is as follows:
DEFault
The format of the trace no. <Chn> (MLOG after *RST); see
CALCulate<Chn>:FORMat
.
MLINear
|z| = sqrt ( x2 + y2 )
MLOGarithmic
MDB
|z| = sqrt ( x2 + y2 )
dB Mag(z) = 20 * log|z| dB
PHASe
W
(z) = arctan ( Im(z) / Re(z) )
POLar
COMPlex
x, y (Real and Imag)
GDELay
Group Delay, – d (z)/d
X
REAL
x
IMAGinary
y
SWR
Standing Wave Ratio, SWR = (1 + |z|) / (1 – |z|)
LINPhase
MLPhase
Lin Mag and Phase, |z|, arctan ( Im(z) / Re(z) )
LOGPhase
MDPhase
dB Mag and Phase, 20 * log|z| dB, arctan ( Im(z) / Re(z) )
IMPedance
R, X, L or C (depending on sign(X))
ADMittance
G, B, L/C (depending on sign(X))