543
SYSTem:ERRor:ALL?
Queries and at the same time deletes all entries in the error queue.
The entries consist of an error number and a short description of the error. Positive error numbers are
instrument-dependent. Negative error numbers are reserved by the SCPI standard; see section Error
Messages.
*RST value
–
SCPI, Command
Types
Confirmed, command or query
Example:
SYST:ERR:ALL?
Query all entries in the error queue. 0,"No error" is returned if the error
queue is empty.
SYSTem:ERRor:CLEar:ALL
Deletes all entries in the table SYSTEM MESSAGES.
*RST value
–
SCPI, Command Types
Device-specific, no query
Example:
SYST:ERR:CLE:ALL
Delete all entries in the table.
SYSTem:ERRor:DISPlay <Boolean>
Switches the display of a tooltip for remote command errors on or off. The tooltip appears at the bottom of
the remote screen and of the manual screen; it is
not
displayed for SCPI errors no. –113,
Undefined
header.
<Boolean>
ON | OFF – Switch the tooltip on or off.
*RST value
OFF
SCPI, Command Types
Device-specific, command or query
Example:
SYST:ERR:DISP ON
Switch the display of the tooltip for remote command errors on.
FREQ:STAR 1
SYSTem:ERRor:LIST?
Reads all system messages and returns a list of comma separated strings. Each string corresponds to an