R&S ZVL
Command Reference
Instrument-Control Commands
Operating Manual 1303.6580.32-05
506
Calculate the system error correction data and apply them to the active channel.
CORR:DATE?
Query the time when the system error correction became active. The analyzer
returns the data and time, e.g.
'03/20/06,18:30:39'
.
CORR:DATA:PAR?
Query the sweep settings for the calibration sweep. The analyzer returns the start
and stop frequency, the number of points, source power, and the sweep type, e.g.
300000,8000000000,201,0,LIN
.
CORR:SST?
Query the calibration status. The analyzer returns
'CAL OFF'
(because the
performed one-port calibration is not sufficient for the measured transmission S-
paramter S
21
).
[SENSe<Ch>:]CORRection:EDELay<port_no>:AUTO ONCE
Defines the offset parameter for the active test port such that the residual delay of the active trace (defined
as the negative derivative of the phase response) is minimized across the entire sweep range.
<Ch>
Channel number of the offset-corrected channel.
<port_no>
Port number of the analyzer. This numeric suffix is ignored; the active port is
determined by the active trace.
ONCE
Applies the
Auto Length
function.
*RST value
–
SCPI, Command
Types
Device-specific, no query
Example:
*RST; CORR:EDEL:AUTO ONCE
Reset the instrument and apply the auto length function to the default trace
(Trc1
in channel 1).
[SENSe<Ch>:]CORRection:EDELay<port_no>:DIELectric <permittivity>
Defines the permittivity for the offset correction at test port <port_no>.
<Ch>
Channel number of the offset-corrected channel
<port_no>
Port number of the analyzer
<permittivity>
Permittivity
Range [def. unit]
1 to +1E+6 []
*RST value
1.00062 [–]
SCPI, Command Types
Device-specific, command or query
Example:
See [SENSe<Ch>:]CORRection:EDELay<port_no>:ELENgth