
R&S ZVL
Command Reference
Instrument-Control Commands
Operating Manual 1303.6580.32-05
555
*RST value
POSitive
SCPI, Command Types
Confirmed, command or query
Example:
TRIG:SOUR EXT
Activate external signal as trigger source.
TRIG:SLOP NEG
Trigger on the negative edge of the (external TTL) trigger signal.
TRIGger<Ch>[:SEQuence]:SOURce IMMediate | EXTernal
Selects the source for the events that the analyzer uses to start a sweep.
<Ch>
Channel number.
IMMediate
Free run measurement without waiting for trigger events
EXTernal
Trigger by external signal applied to the
EXT TRIGGER / GATE IN
connector
on the rear panel.
*RST value
IMMediate
SCPI, Command
Types
Confirmed, command or query.
Example:
TRIG:SOUR EXT
Activate external trigger mode.
UNIT...
This subsystem defines default units for all SCPI commands and queries implemented by the network
analyzer.
UNIT:LENGth <unit>
Defines the default unit of length. The command affects all SCPI commands defining a parameter with a
physical unit that is related to a length, in particular:
Length offset parameters (
[SENSe:]CORRection:EDELay...
)
Parameters of the distance-to fault measurement
(
CALCulate<Chn>:TRANSform:DTFault...
)
Derived units, e.g. the attenuation per unit of length defined via
CALCulate<Ch>:TRANsform:DTFault:DEFine
Stimulus values for limit lines and ripple limit lines for the distance-to fault trace (e.g.
CALCulate<Chn>:LIMit:DATA
)
<unit>
METer– metric units
FOOT – foot (ft). 1 foot corresponds to 0.3048 m (1 m = 3.28084 ft).