R&S ZVL
Command Reference
Instrument-Control Commands
Operating Manual 1303.6580.32-05
474
MMEMory:STORe:CABLe '<target_directory>'
Copies all pre-defined and user-defined cable description files (*.rsc) to a specified directory. The contents
of the source directories for cable description files (
C:\R_S\Instr\resources\Nwa\CableTypes
for
predefined,
C:\R_S\Instr\user\Nwa\CableTypes
for user-defined cables) are not affected.
'<target_directory>'
String parameter to specify the target directory.
*RST value
–
SCPI, Command
Types
Device-specific, no query.
Example:
MMEM:STOR:CABL 'c:\temp'
Copy all *rsc files from the source directories (see
above) to C:\temp.
After a possible modification of the *.rsc files in
c:\temp:
MMEM:LOAD:CABL 'c:\temp'
Re-load all *rsc files from C:\temp. The files remain
in c:temp; they are not copied to the default
directories for cable description files.
MMEMory:STORe:CKIT '<kit_name>', '<file_name>'
Stores the data of a calibration kit to a specified file.
'<kit_name>'
Name of a user-defined calibration kit available on the analyzer.
It is not possible to modify or store predefined or ideal kits.
'<file_name>'
String parameter to specify the name and directory of the cal kit file to be created.
The file is a NWA-specific cal kit file with the extension *.calkit. If no path is specified
the analyzer uses the current directory, to be queried with
MMEMory:CDIRectory?
.
*RST value
–
SCPI,
Command
Types
Device-specific, no query.
Example:
MMEM:LOAD:CKIT
'C:\R_S\Instr\user\Nwa\Calibration\Kits\New_kit.calkit'
Load the previously created cal kit file New_kit.calkit from
the default cal kit directory.
... :MMEM:STOR:CKIT 'New_kit',
'C:\R_S\Instr\user\Nwa\Calibration\Kits\New_kit.calkit'
Store the data for the user-defined cal kit Newkit and
overwrite the cal kit file New_kit.calkit.