<Ch>
Channel number of the calibrated channel (not relevant for this
command).
*RST value
–
SCPI,
Command
Types
Device-specific, query only
Example:
CORR:COLL:AUTO:PORT '', 1, 2, 2, 1
Perform an automatic 2-port TOSM calibration at the analyzer test ports
1 and 2 using the calibration unit's default calibration kit file and ports 2
and 1 of the cal unit.
CORR:COLL:AUTO:PORT:CONN?
Query the actual port assignment. If the cal unit is properly connected
according to the previous command, the response is
1,2,2,1.
A zero
would mean that the corresponding analyzer port is not connected to any
port of the calibration unit.
[SENSe<Ch>:]CORRection:COLLect:CONNection<port_no> N50FEMALE |
N50MALE | N75FEMALE | N75MALE | PC7 | PC35FEMALE | PC35MALE |
PC292FEMALE | PC292MALE
Additional parameters (): UFEMALE1 | UMALE1 | UFEMALE2 | UMALE2
Selects a connector type at a specified port <port_no> and its gender.
Use
[SENSe<Ch>:]CORRection:COLLect:SCONnection<port_no>
to select an arbitrary
connector type using a string variable.
<Ch>
Channel number of the calibrated channel.
<port_no>
Port numbers of the analyzer.
Note:
If the analyzer is set to use the same connectors at all ports
(
[SENSe<Ch>:]CORRection:COLLect:CONNection<port_no>:PORTs
ALL
), then a change of a connector type is valid for all ports. The gender of the
connectors can still be different.
Parameters
Connector type and gender of the connectors (omitted for query). UFEMALE1 and
UMALE1 denote the user-defined connector type
UserConn1,
UFEMALE2 and
UMALE2 denote the user-defined connector type
UserConn2.
The user-defined
connector types must be defined before being addressed by
[SENSe<Ch>:]CORRection:COLLect:CONNection<port_no>.
*RST value
N50FEMALE for all ports.
SCPI,
Command
Types
Device-specific, command or query.
Example:
*RST; CORR:COLL:CONN1 N75MALE; CONN4?
Change the connector type at port 1 from
N50FEMALE
to
N75MALE
. The connector
type at the other ports is also changed to N75, however, the gender (female) is