R&S ZVL
Command Reference
Instrument-Control Commands
Operating Manual 1303.6580.32-05
545
221, (settings conflict).
*RST value
– (neither the password nor the protection of service functions is affected by
*RST)
SCPI, Command
Types
Confirmed, no query
Example:
SYST:PASS "XXXX"
Enter password.
SYSTem:PASSword:RESet
This command resets the service password.
*RST value
–
SCPI, Command Types
Confirmed, no query
Example:
SYST:PASS:RES
Reset password.
SYSTem:PRESet
Performs a factory preset of all instrument settings (i.e. all open setups) or of the active setup, depending
on the
SYSTem:PRESet:SCOPe
settings. The command is equivalent to *RST and to the action of the
PRESET
key on the front panel.
If a user-defined preset is active (
SYSTem:PRESet:USER[:STATe] ON
), the PRESET key initiates a
user-defined preset, while
SYSTem:PRESet
and
*RST
still activate the factory preset.
*RST value
–
SCPI, Command Types
Confirmed, no query
Example:
SYST:PRES:SCOP SING
Define the scope of a preset: the active setup is reset only.
SYST:PRES
Reset the parameters of the current setup.
SYSTem:PRESet:SCOPe ALL | SINGle
Specifies whether a preset (SYSTem:PRESet;
*RST
) affects the active setup only or all open setups.
This command is effective in the Network Analyzer mode only and returns an error in other modes.
ALL
All open setups are deleted and the setup
Set1
is created with default trace and
channel settings.