98
ENGLISH
Select “Menu”
→
“CO
2
” to enter into CO
2
related setting.
In RESP settings window, CO
2
Settings can be entered if your monitor is configured with CO
2
monitoring .
:
click it to
turn on or off CO
2
Monitoring. When the CO
2
monitoring is selected, then all items
setting as “RESP” will shift to “CO
2
”.
See Chapter Monitoring Carbon Dioxide (CO
2
)
.
On RESP Settings screen, there is CO
2
Monitoring Switch
:
click it to
turn on or off CO
2
Monitoring.
It is recommended that the switch is turned on only when there is a need to monitor CO
2
parameter. This can not only
reduce the power consumption and also extend the life of the CO
2
measurement module.
In order to maintain the monitor, please set CO
2
switch at OFF state in system setup when CO
2
function is not used.
Gain:
the CO
2
waveform gain. 4 options: X1/2, X1, X2 and X4. The default is X1 for adult and pediatric patient, and
X2 for neonate patient.
X1 waveform scale with base gain X1/2 half scale size of the base gain
X2 twice scale size of the base gain X4 four times scale size of the base gain
Speed:
Respiration waveform sweeping speed of, 2 options 6.25mm/s and 12.5 mm/s. The default is 12.5 mm/s.
Apnea(s):
The timeout setting for apnea alarm (in second).
(1). When CO
2
monitoring is on:
It can be set as any number from 10 to 60 seconds, the step is 1 second. Icon “
“ displays on the
upper right corner of Respiration Panel; When the device has not detected any breathing signal for the specified
time, the “Apnea” alarm prompts on, and alarm sound will be activated. If it is set as OFF, the icon “
“
displays on lower left corner of the Respiration Panel. The default is 20s.
(2). When CO
2
monitoring is off:
refer to Section
Changing RESP setting
.
Source:
Respiration signal source. This item is fixed to be “CO
2
” if CO
2
monitoring function is selected. Otherwise, the
source will be obtained from thoracic impedance measurement from ECG module.
Zero:
press it to perform zero reset.
Default:
resume to the factory default value.
Unit:
It can be set up as “%”, “kPa” and “mmHg”. If the unit is changed, then the parameter value will change and
refresh timely. The unit will be displayed on parameter area, the default is “mmHg”.
Period:
setting the calculating cycle of the EtCO
2
value, there are three selectable options: “1b”, “10s” and “20s”.
The default is “10s”. “1b” means the EtCO
2
value will be calculated once every respiration cycle; “10s” means the
EtCO
2
value will be calculated once every 10 seconds, and the maximum EtCO
2
value measured during this 10s will
be displayed on data area; “20s” means the EtCO
2
value will be calculated once every 20 seconds, and the maximum
EtCO
2
value measured during this 20s will be displayed on data area.
Balance:
setting the balance gas in patient’s respiration air flow. There are three kinds of selectable balance gas:
“Air”, “N
2
O” and “He”, namely: air, nitrous oxide and helium. If no specific balance gas is given, the balance gas can
be set as “Air”.
O
2
Comp.:
adjusting the concentration of the compensating gas in patient’s respiration air flow. Generally, the
compensating gas is Oxygen, so it can be called oxygen compensation concentration. The unit: %; Setting range:
Содержание 35306
Страница 2: ...2 ENGLISH Patient Monitor Operator s Manual ...
Страница 21: ...21 ENGLISH Series III Front view Rear view Left view Right view 11 ...
Страница 22: ...22 ENGLISH Series IV Front view Rear view Left view Right view 11 Extended Module ...
Страница 128: ...128 ENGLISH Oxygenation calculation Calculate result Oxygenation calculation Reference range Unit Result Range ...
Страница 171: ...2 ESPAÑOL Monitor de paciente Manual de operador ...
Страница 189: ...20 ESPAÑOL Serie III Vista delantera Vista posterior Vista lateral izquierda Vista lateral derecha 11 ...
Страница 190: ...21 ESPAÑOL Serie IV Vista delantera Vista posterior Vista lateral izquierda Vista lateral derecha 11 Módulo adicional ...
Страница 343: ...2 FRANÇAIS Moniteur patient Manuel d utilisation ...
Страница 361: ...20 FRANÇAIS Série II Vue de face Vue arrière Connecteur et icône Vue de gauche Vue de droite 13 12 ...
Страница 362: ...21 FRANÇAIS Série III Vue de face Vue arrière Vue de gauche Vue de droite 11 ...
Страница 363: ...22 FRANÇAIS Série IV Vue de face Vue arrière Vue de gauche Vue de droite Panneau avant 11 Module avancé ...
Страница 519: ...2 ITALIANO Monitor Paziente Manuale d Uso ...
Страница 538: ...21 ITALIANO Serie III Vista anteriore Vista posteriore Lato sinistro Lato destro 11 ...
Страница 539: ...22 ITALIANO Serie IV Vista anteriore Vista posteriore Lato sinistro Lato destro 11 Modulo esteso ...