76
ENGLISH
1. Power on the monitor.
2. Check the patient information area on the screen. Set a correct patient type, select a correct cuff size.
3. Connect the tube with cuff to the connector marked with “NIBP” icon on the signal input panel.
4. Select a cuff with correct size, then unfold the cuff and wrap it around the patient’s upper arm as follows:
Determine the patient’s limb circumference.
Select an appropriate cuff by referring to the limb circumference marked on the cuff. The width of the cuff
should be 40% of the limb circumference, or 2/3 of the upper arm’s length. The inflatable part of the cuff should
be long enough to encircle at least 50% to 80% of the limb. When putting on the cuff, unfold and wrap it around
the upper arm evenly to appropriate tightness.
Remember to empty the residual air in the cuff before the measurement is commenced.
Locate the cuff in such a way that the artery mark “” is at a location where the clearest pulsation of brachial
artery is observed.
The cuff should be tightened to a degree where insertion of one finger is allowed.
The lower end of the cuff should be 2cm above the elbow joint.
9.5.2 Starting and Stopping Measurements
Starting and Stopping NIBP Measurements by pressing NIBP measure key “
“.
Too frequent blood pressure measurements may cause purpura, ischemia and neuropathy in the limb with the cuff.
Inspect the application site regularly to ensure skin quality and inspect the extremity of the cuffed limb for normal color,
warmth and sensitivity. If any abnormity occurs, move the cuff to another site or stop the blood pressure measurements
immediately.
Note
: You can not start NIBP measurement if the NIBP parameter area is not displayed.
9.5.3 Factors Affecting NIBP Measurement
Like common non-invasive blood pressure measurement, improper operation may cause inaccurate or blank result
or misunderstanding of the measuring information when the oscillometric method is used to take the blood pressure
measurement. These points need particular attention of the operators.
Содержание 35306
Страница 2: ...2 ENGLISH Patient Monitor Operator s Manual ...
Страница 21: ...21 ENGLISH Series III Front view Rear view Left view Right view 11 ...
Страница 22: ...22 ENGLISH Series IV Front view Rear view Left view Right view 11 Extended Module ...
Страница 128: ...128 ENGLISH Oxygenation calculation Calculate result Oxygenation calculation Reference range Unit Result Range ...
Страница 171: ...2 ESPAÑOL Monitor de paciente Manual de operador ...
Страница 189: ...20 ESPAÑOL Serie III Vista delantera Vista posterior Vista lateral izquierda Vista lateral derecha 11 ...
Страница 190: ...21 ESPAÑOL Serie IV Vista delantera Vista posterior Vista lateral izquierda Vista lateral derecha 11 Módulo adicional ...
Страница 343: ...2 FRANÇAIS Moniteur patient Manuel d utilisation ...
Страница 361: ...20 FRANÇAIS Série II Vue de face Vue arrière Connecteur et icône Vue de gauche Vue de droite 13 12 ...
Страница 362: ...21 FRANÇAIS Série III Vue de face Vue arrière Vue de gauche Vue de droite 11 ...
Страница 363: ...22 FRANÇAIS Série IV Vue de face Vue arrière Vue de gauche Vue de droite Panneau avant 11 Module avancé ...
Страница 519: ...2 ITALIANO Monitor Paziente Manuale d Uso ...
Страница 538: ...21 ITALIANO Serie III Vista anteriore Vista posteriore Lato sinistro Lato destro 11 ...
Страница 539: ...22 ITALIANO Serie IV Vista anteriore Vista posteriore Lato sinistro Lato destro 11 Modulo esteso ...