Wave 1 can be set as any ECG lead, or turn off this channel directly.
Wave 2 and Wave 3 can be set as OFF, another ECG wave different from that selected in Wave I, SpO
2
, and RESP
(or CO
2
). If “3-lead wires” is selected, then Wave 2 and Wave 3 can be set as SpO
2
, and RESP (or CO
2
), or set as the
current cascaded ECG waveform (e.g. ECGII).
Parameter Settings:
“Param 1”, “Param 2” and “Param 4” can be set as OFF, ECG, SpO
2,
NIBP, TEMP and RESP (or CO
2
) and so on.
“Param 3” can be set as ECG, TEMP, NIBP, SpO
2
and
RESP (or CO
2
) and so on.
Cascaded above: select it means all first 3 channels waveform display the waveform set in Waveform 1, and
waveform 2 and waveform 3 are the cascaded waveform of Waveform 1. And at the moment, Wave 2 and Wave 3
are nonadjustable. The default is deselect it.
Close all waveforms: select it means the waveform 1 to waveform 3 will not be displayed on the waveform panel in
the big font view, but the information in parameter panel will be displayed.
5.2.3 All ECG Trace view
1) Description:
In this View, all ECG trace waveforms are displayed on screen simultaneously. All 7 traces of ECG waveform are displayed
on the left, and the 5 parameter panels are on the right.
2) All ECG Trace View Settings:
The default 12 waveforms displaying in the view are: ECG I, ECG II, ECG III, ECG aVR, ECG aVL, ECG aVF, ECG V1, ECG V2,
ECG V3, ECG V4, ECG V5, ECG V6, waveform 1~12 are nonadjustable.
The default 7 waveforms displaying in the view are: ECG I, ECG II, ECG III, ECG aVR, ECG aVL, ECG aVF and ECG V, waveform
1~7 are nonadjustable.
When “3-lead wires” is selected, then it displays the cascaded waveform, the cascaded waveform is fixed to be “ECG II”,
“ECG I” or “ECG III”.
Parameter 1 and Parameter 2 are nonadjustable, the default settings are ECG and SpO
2
respectively.
Parameter 3 to Parameter 5 can be set as OFF, TEMP, NIBP, RESP (or CO
2
) and so on.
All waves:
select this item, then all waveforms (including all ECG waveforms and other waveforms) will be displayed
on ECG trace view. The factory default is un-select.
5.2.4 NIBP List view
1) Description:
In this View, 4 waveform channels are default to be ECG I, ECG II, ECG III and SpO
2
plethysmogram. Parameter panels in
the parameter area are default to ECG, SpO
2
, NIBP, RESP and TEMP. At the same view, the NIBP list view will be displayed.
Содержание 35306
Страница 2: ...2 ENGLISH Patient Monitor Operator s Manual ...
Страница 21: ...21 ENGLISH Series III Front view Rear view Left view Right view 11 ...
Страница 22: ...22 ENGLISH Series IV Front view Rear view Left view Right view 11 Extended Module ...
Страница 128: ...128 ENGLISH Oxygenation calculation Calculate result Oxygenation calculation Reference range Unit Result Range ...
Страница 171: ...2 ESPAÑOL Monitor de paciente Manual de operador ...
Страница 189: ...20 ESPAÑOL Serie III Vista delantera Vista posterior Vista lateral izquierda Vista lateral derecha 11 ...
Страница 190: ...21 ESPAÑOL Serie IV Vista delantera Vista posterior Vista lateral izquierda Vista lateral derecha 11 Módulo adicional ...
Страница 343: ...2 FRANÇAIS Moniteur patient Manuel d utilisation ...
Страница 361: ...20 FRANÇAIS Série II Vue de face Vue arrière Connecteur et icône Vue de gauche Vue de droite 13 12 ...
Страница 362: ...21 FRANÇAIS Série III Vue de face Vue arrière Vue de gauche Vue de droite 11 ...
Страница 363: ...22 FRANÇAIS Série IV Vue de face Vue arrière Vue de gauche Vue de droite Panneau avant 11 Module avancé ...
Страница 519: ...2 ITALIANO Monitor Paziente Manuale d Uso ...
Страница 538: ...21 ITALIANO Serie III Vista anteriore Vista posteriore Lato sinistro Lato destro 11 ...
Страница 539: ...22 ITALIANO Serie IV Vista anteriore Vista posteriore Lato sinistro Lato destro 11 Modulo esteso ...