163
ENGLISH
B Status/Error during ECG, SpO
2
, NIBP Monitoring
“Cuff error”
cuff is not wrapped correctly, or is not connected
“Air leak”
Air moving part, tube or the cuff leak air.
“Pressure error”
Unstable cuff pressure or tangled cuff tubing
“Signal weak”
Very weak signal because of the cuff, or the patient has very weak pulse
“Over extent”
The measurement range exceeds 255 mmHg (Pediatric patient over 135
mmHg)
“Over motion”
The repeated measurement due to moving, excessive noise during the
stepping inflation and measuring pressure and pulse, e.g. during patient
shaking motion
“Signal overflow”
Blood pressure amplifier overflow due to excessive movement
“Leak in gas run”
Leaking during the pneumatic device testing
“System error”
Abnormal condition of CPU, such as register overflow, divided by zero
“Adult”
The blood pressure measuring now is in adult mode. In this case, it is not
allowed to monitoring Pediatric or neonatal patient. Otherwise, there may be
serious danger to the Pediatric monitored.
“Pediatric”
The blood pressure module is now worked in Pediatric measuring mode.
“PROBE OFF”
SpO
2
probe fell off
“LEADS OFF”
The ECG electrodes or cable fell off
“LEARNING”
Learning arrhythmia for 15 seconds
“DEMO”
The monitor is displaying the demo waveforms, which are generated by the
monitor itself.
C Status/Error during CO
2
Monitoring
Suggested Message/Response
Description
“Sensor Over Temp”
Make sure sensor is not exposed to extreme heat (heat
lamp, etc.). If error persists, return sensor to factory for
servicing.
The sensor temperature is greater than 40 °C.
“Sensor Faulty”
Check that the sensor is properly plugged in. Reinsert
or reset the sensor if necessary. If error persists, return
sensor to factory for servicing
One of the following conditions exist:
Source Current Failure,
EEPROM Checksum Faulty,
Hardware Error
No Parameter Message
The host must set the Barometric Pressure and
compensations to clear this error; no user intervention
should be required.
Barometric Pressure and/or gas compensations have not
been set since power on. For CO
2
to be calculated with
the stated accuracy, these values should be set whenever
the sensor is plugged in.
“Module in Sleep Mode”
This bit is set when sensor has been placed in sleep
mode.
Содержание 35306
Страница 2: ...2 ENGLISH Patient Monitor Operator s Manual ...
Страница 21: ...21 ENGLISH Series III Front view Rear view Left view Right view 11 ...
Страница 22: ...22 ENGLISH Series IV Front view Rear view Left view Right view 11 Extended Module ...
Страница 128: ...128 ENGLISH Oxygenation calculation Calculate result Oxygenation calculation Reference range Unit Result Range ...
Страница 171: ...2 ESPAÑOL Monitor de paciente Manual de operador ...
Страница 189: ...20 ESPAÑOL Serie III Vista delantera Vista posterior Vista lateral izquierda Vista lateral derecha 11 ...
Страница 190: ...21 ESPAÑOL Serie IV Vista delantera Vista posterior Vista lateral izquierda Vista lateral derecha 11 Módulo adicional ...
Страница 343: ...2 FRANÇAIS Moniteur patient Manuel d utilisation ...
Страница 361: ...20 FRANÇAIS Série II Vue de face Vue arrière Connecteur et icône Vue de gauche Vue de droite 13 12 ...
Страница 362: ...21 FRANÇAIS Série III Vue de face Vue arrière Vue de gauche Vue de droite 11 ...
Страница 363: ...22 FRANÇAIS Série IV Vue de face Vue arrière Vue de gauche Vue de droite Panneau avant 11 Module avancé ...
Страница 519: ...2 ITALIANO Monitor Paziente Manuale d Uso ...
Страница 538: ...21 ITALIANO Serie III Vista anteriore Vista posteriore Lato sinistro Lato destro 11 ...
Страница 539: ...22 ITALIANO Serie IV Vista anteriore Vista posteriore Lato sinistro Lato destro 11 Modulo esteso ...