110
ENGLISH
2. Plug the Nomoline sampling line into the ISA input port.
3. Connect the gas outlet with the releasing system, or make the gas flow to the patient circuit.
4. Power on the monitor.
5. A green LED indicates that the ISA gas analyzer is ready for use.
6. Do pre-use check following the section part.
2. Pre-use Check
Before plugging the Nomoline sampling line into the respiratory circuit, please do the following operation:
1. Connect the sampling line to the LEGI.
2. Check whether the green light in LEGI is normal.
3. For ISA OR+ and ISA AX+: Check whether the gas (O
2
) reading is correct (21%).
4. Expiring gas towards the sampling line and check that the gas reading (CO
2
) and waveform are displayed on the monitor.
5. Clogging the sampling line with finger for 10 seconds.
6. Check if the clogging alarm is displayed and red light in LEGI is blinking.
7. Check whether the patient circuit is hermetic.
3. LEGI Status
Color of LED
Information
Steady green light
System OK
Blinking green light
Zeroing in process
Steady blue light
Anesthetic agent present
Blinking red light
Check sampling line
Steady red light
Sensor error
4. Zero Procedure
Infrared gas analyzer requires to ensure the zero datum level of the monitoring data of CO2, N2O and the anesthetic
agents (AA). This zero calibration is called Zeroing.
ISA sidestream analyzer is intended to perform Self-Zeroing by shifting the sampling gas from breathing circuit to ambient
environment. The device performs Zeroing once a day, and approximately 10 seconds while Zeroing is in progress.
5. Instruction for Safe Operation --- ISA sidestream
ISA sidestream analyzer is intended for use by authorized and trained medical personnel only.
Use only PHASEIN manufactured Nomoline sampling line.
Содержание 35306
Страница 2: ...2 ENGLISH Patient Monitor Operator s Manual ...
Страница 21: ...21 ENGLISH Series III Front view Rear view Left view Right view 11 ...
Страница 22: ...22 ENGLISH Series IV Front view Rear view Left view Right view 11 Extended Module ...
Страница 128: ...128 ENGLISH Oxygenation calculation Calculate result Oxygenation calculation Reference range Unit Result Range ...
Страница 171: ...2 ESPAÑOL Monitor de paciente Manual de operador ...
Страница 189: ...20 ESPAÑOL Serie III Vista delantera Vista posterior Vista lateral izquierda Vista lateral derecha 11 ...
Страница 190: ...21 ESPAÑOL Serie IV Vista delantera Vista posterior Vista lateral izquierda Vista lateral derecha 11 Módulo adicional ...
Страница 343: ...2 FRANÇAIS Moniteur patient Manuel d utilisation ...
Страница 361: ...20 FRANÇAIS Série II Vue de face Vue arrière Connecteur et icône Vue de gauche Vue de droite 13 12 ...
Страница 362: ...21 FRANÇAIS Série III Vue de face Vue arrière Vue de gauche Vue de droite 11 ...
Страница 363: ...22 FRANÇAIS Série IV Vue de face Vue arrière Vue de gauche Vue de droite Panneau avant 11 Module avancé ...
Страница 519: ...2 ITALIANO Monitor Paziente Manuale d Uso ...
Страница 538: ...21 ITALIANO Serie III Vista anteriore Vista posteriore Lato sinistro Lato destro 11 ...
Страница 539: ...22 ITALIANO Serie IV Vista anteriore Vista posteriore Lato sinistro Lato destro 11 Modulo esteso ...