Please follow the corresponding methods to make temperature measurement according to the temperature transducer
you selected.
Connecting the thermal temperature sensor:
The temperature sensor is thermo-resistor type it needs time to respond the temperature change, so the accurate
temperature value displays after a while. The temperature sensor equipped may have different shape for measuring body
surface temperature or cavity temperature respectively.
Normal value for body surface: 36.5°C ~37°C;
Notes:
For body surface temperature probe, attach the TEMP sensor to the patient, generally if the TEMP sensor and skin
doesn’t contact closely, the measured value becomes lower, so for those who have requirement for temperature
monitoring, add a proper pad to the sensor and fix it with adhesive tape to make them contact firmly.
Especially for pediatric patient, they have more activities, pay more attention to the sensor fixing.
The TEMP sensor is designed for use with the specific patient monitor, which cannot be used as applied part to
other products.
The operator is responsible for checking the compatibility of the patient monitor and sensor type including cable
before use.
Incompatible components can result in degraded performance.
Operation Procedures for thermal temperature transducer:
1. Securely attach the transducer to the patient;
2. Connect the cable to TEMP probe connector marked “TEMP” in the panel.
3. Check that the menu setting is matching the used temperature sensor type.
4. Check that the alarm settings are appropriate for this patient.
Note: When unplugging the probe, be sure to hold the head of the connector and pull it out.
11.4 Understanding the TEMP Display
1. TEMP1
℃
: Temperature 1 mark and unit. Temperature unit:
℃
or
℉。
2. 36.5: displays the temperature measured at temperature channel 1.
3. TEMP 2: Temperature channel 2.
4. 37.0: displays the temperature measured at temperature channel 2.
1
3
4
2
5
Содержание 35306
Страница 2: ...2 ENGLISH Patient Monitor Operator s Manual ...
Страница 21: ...21 ENGLISH Series III Front view Rear view Left view Right view 11 ...
Страница 22: ...22 ENGLISH Series IV Front view Rear view Left view Right view 11 Extended Module ...
Страница 128: ...128 ENGLISH Oxygenation calculation Calculate result Oxygenation calculation Reference range Unit Result Range ...
Страница 171: ...2 ESPAÑOL Monitor de paciente Manual de operador ...
Страница 189: ...20 ESPAÑOL Serie III Vista delantera Vista posterior Vista lateral izquierda Vista lateral derecha 11 ...
Страница 190: ...21 ESPAÑOL Serie IV Vista delantera Vista posterior Vista lateral izquierda Vista lateral derecha 11 Módulo adicional ...
Страница 343: ...2 FRANÇAIS Moniteur patient Manuel d utilisation ...
Страница 361: ...20 FRANÇAIS Série II Vue de face Vue arrière Connecteur et icône Vue de gauche Vue de droite 13 12 ...
Страница 362: ...21 FRANÇAIS Série III Vue de face Vue arrière Vue de gauche Vue de droite 11 ...
Страница 363: ...22 FRANÇAIS Série IV Vue de face Vue arrière Vue de gauche Vue de droite Panneau avant 11 Module avancé ...
Страница 519: ...2 ITALIANO Monitor Paziente Manuale d Uso ...
Страница 538: ...21 ITALIANO Serie III Vista anteriore Vista posteriore Lato sinistro Lato destro 11 ...
Страница 539: ...22 ITALIANO Serie IV Vista anteriore Vista posteriore Lato sinistro Lato destro 11 Modulo esteso ...