43
ENGLISH
Wave 3:
print the third channel of waveform.
Note:
The setting for Wave 1, Wave 2 and Wave 3 should be different (eg. Wave 2 and Wave 3 can not be set as “ECG
III” at the same time). Only 2 ECG waveforms at most can be printed simultaneously.
ARR Trigger:
ARR triggered printing. “ON” means the printer will print out waveform once an ARR event occurs. The
default is “OFF”.
Duration:
the duration of printing real-time waveform. Options are: 10, 20, 30, 40, 50, 60 (seconds), and Continuous,
the default is “10” seconds. If “Continuous” is selected, then the printer will not stop printing until manual stop it by
pressing the print key.
Print ECG Grids:
the switch for grids on printed ECG waveform. Select it means the 5x5(mm
2
) grid background will
be printed when printing the ECG waveform.
Timer:
the switch for print timer. The default is “OFF”. If Print is ON, rotate navigation knob to set on the Timer to
enable timed print, and set the value of printing intervals in the cycle category. The setting range is 5~480 minutes,
the step is 5 minutes. The default is 60 minutes.
CSM (optional):
select the CSM parameter, options are “EEG” and “Trend” that means the corresponding
electroencephalogram and graphic trend will be printed respectively.
3.9 Operating Modes
3.9.1 Real Time Mode
The “Real Time” shows the normal work mode with real time waveform and data.
3.9.2 Demo Mode
The “Demo” mode shows the demo working mode with simulated waveform and data. In the “Demo” mode, all the
signals and data are generated by the device for demonstrating and test purpose (the alarm function can be checked in
this mode).
The Demo mode is for demonstration purpose only. To avoid stimulated data being mistaken for a monitored
patient’s data, you must not change into Demo mode during monitoring. Otherwise, improper monitoring and
delayed treatment could result.
3.9.3 Night Mode
You can select night mode to avoid disturbing the patient during monitoring. In the night mode, the pulse beep and key
tone are off, alarm volume is set to level 2, and brightness of the display decreases to level 1 by default.
Verify the alarm volume and pulse beep volume settings after entering night mode. You can adjust the settings
according to your requirements. Pay attention to the potential risk if the setting value is too low.
To enter night mode, select “Menu”
→
”System Settings”
→
“General”
→
“Mode”
→
“Night Mode”.
To exit night mode, select the other desirable mode in “Mode”.
Содержание 35306
Страница 2: ...2 ENGLISH Patient Monitor Operator s Manual ...
Страница 21: ...21 ENGLISH Series III Front view Rear view Left view Right view 11 ...
Страница 22: ...22 ENGLISH Series IV Front view Rear view Left view Right view 11 Extended Module ...
Страница 128: ...128 ENGLISH Oxygenation calculation Calculate result Oxygenation calculation Reference range Unit Result Range ...
Страница 171: ...2 ESPAÑOL Monitor de paciente Manual de operador ...
Страница 189: ...20 ESPAÑOL Serie III Vista delantera Vista posterior Vista lateral izquierda Vista lateral derecha 11 ...
Страница 190: ...21 ESPAÑOL Serie IV Vista delantera Vista posterior Vista lateral izquierda Vista lateral derecha 11 Módulo adicional ...
Страница 343: ...2 FRANÇAIS Moniteur patient Manuel d utilisation ...
Страница 361: ...20 FRANÇAIS Série II Vue de face Vue arrière Connecteur et icône Vue de gauche Vue de droite 13 12 ...
Страница 362: ...21 FRANÇAIS Série III Vue de face Vue arrière Vue de gauche Vue de droite 11 ...
Страница 363: ...22 FRANÇAIS Série IV Vue de face Vue arrière Vue de gauche Vue de droite Panneau avant 11 Module avancé ...
Страница 519: ...2 ITALIANO Monitor Paziente Manuale d Uso ...
Страница 538: ...21 ITALIANO Serie III Vista anteriore Vista posteriore Lato sinistro Lato destro 11 ...
Страница 539: ...22 ITALIANO Serie IV Vista anteriore Vista posteriore Lato sinistro Lato destro 11 Modulo esteso ...