111
ENGLISH
The ISA sidestream analyzer must not be used with flammable anesthetic agents.
The sampling line should be kept away from the patient’s cervix to prevent any possible suffocation of the patient.
Reusing the one-off sampling line is forbidden.
The used sampling line shall be disposed of in accordance with local regulations for medical waste.
Do not use the Adult/Pediatric sampling line with infants as the adapter adds dead space to the patient circuit.
Do not use the Infant sampling line with adult as this may cause excessive flow resistance.
Do not use the ISA sidestream analyzer together with the metered dose inhaler propellants as this may cause the
bacilli strainer clogged.
Zeroing success needs the gas analyzer existing in the ambient air (21% O2 and 0% CO2). Therefore, please place the
ISA on draughty position. During performing zeroing procedure, please avoid breathing around the ISA sidestream
analyzer.
Nomoline sampling line and its interface are non-sterile device. Do not autoclave the devices as this will damage it.
Never sterilize or immerse the ISA sidestream analyzer.
The portable and mobile RF communications equipment will affect the measurement. Please make sure that the ISA
sidestream analyzer is used in the specified electromagnetic environment.
ISA sidestream analyzer is intended to be used as a means of auxiliary device of monitoring a patient. It shall always
be used in combination with other Patient Monitor.
Please change the sampling line if red light is blinking on LEGI or “Sampling line clogged” displays on the monitor.
ISA sidestream analyzer is not intended to be used in MRI environment.
Placing the device out of the MRI Room during MRI scanning.
Unprecise measured result may cause if High frequency electro-surgical device is used around the ISA or monitor.
Do not operate the external self-cool function of the ISA device.
Do not eliminate the condensate by adding negative pressure (eg. Use injector to add negative pressure) to Nomoline.
The overed positive pressure or negative pressure inpatient’s circuit will affect the flowing speed.
To prevent the Analyzer from dropping on the patient, the ISA sidestream analyzer should be well placed.
Do not apply tension to the cable of the ISA sidestream analyzer.
Please fix the ISA “Plug and Play” analyzer to avoid causing damage on ISA sidestream analyzer.
Do not operate the ISA sidestream analyzer outside the specified operating temperature environment.
Содержание 35306
Страница 2: ...2 ENGLISH Patient Monitor Operator s Manual ...
Страница 21: ...21 ENGLISH Series III Front view Rear view Left view Right view 11 ...
Страница 22: ...22 ENGLISH Series IV Front view Rear view Left view Right view 11 Extended Module ...
Страница 128: ...128 ENGLISH Oxygenation calculation Calculate result Oxygenation calculation Reference range Unit Result Range ...
Страница 171: ...2 ESPAÑOL Monitor de paciente Manual de operador ...
Страница 189: ...20 ESPAÑOL Serie III Vista delantera Vista posterior Vista lateral izquierda Vista lateral derecha 11 ...
Страница 190: ...21 ESPAÑOL Serie IV Vista delantera Vista posterior Vista lateral izquierda Vista lateral derecha 11 Módulo adicional ...
Страница 343: ...2 FRANÇAIS Moniteur patient Manuel d utilisation ...
Страница 361: ...20 FRANÇAIS Série II Vue de face Vue arrière Connecteur et icône Vue de gauche Vue de droite 13 12 ...
Страница 362: ...21 FRANÇAIS Série III Vue de face Vue arrière Vue de gauche Vue de droite 11 ...
Страница 363: ...22 FRANÇAIS Série IV Vue de face Vue arrière Vue de gauche Vue de droite Panneau avant 11 Module avancé ...
Страница 519: ...2 ITALIANO Monitor Paziente Manuale d Uso ...
Страница 538: ...21 ITALIANO Serie III Vista anteriore Vista posteriore Lato sinistro Lato destro 11 ...
Страница 539: ...22 ITALIANO Serie IV Vista anteriore Vista posteriore Lato sinistro Lato destro 11 Modulo esteso ...