![Gima 35306 Скачать руководство пользователя страница 157](http://html1.mh-extra.com/html/gima/35306/35306_manual_2228144157.webp)
157
ENGLISH
Electromagnetic Immunity
The IRMA probe is intended for use in the electromagnetic environment specified below. The customer or the user of
the IRMA probe should assure that it is used in such an environment.
Immunity test
IEC 61001 test level
Compliance
level
Electromagnetic environment- guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
±
6 kV contract
±
6 kV air
±
6 kV
contract
±
6 kV air
Floor should be wood, concrete or ceramic tile. If
floors are covered with synthetic material, the relative
humidity should be at least 30%.
Power frequency
(50/60Hz) magnetic
field
IEC 61000-4-8
3A/m
3A/m
Mains power quality should be that of a typical
commercial or hospital environment.
RF emissions
GISPR 11
N/A
Group 1
Mains power quality should be that of a typical
commercial or hospital environment. If the use of
the IRMA probe requires continued operation during
power mains interruptions, it is recommended that
the IRMA probe is powered from an uniterruptible
power supply or a battery
RF emissions
GISPR 11
N/A
Group B
The IRMA probe uses RF energy only for its internal
function. Therefore, its RF emissions are very low
and are not likely to cause any interference in nearby
electronic equipment.
Conducted RF
IEC 61000-4-6
3 Vrms
150 kHz - 180 MHz
10 Vrms
Portable and mobile RF communications equipment
should be used no closer to any part of ISA sidestream
analyzer, including cables, than the recommended
separation distance calculated from the equation
applicable to the frequency of the transmitter.
Recommended separation distance
d=0.35
P
d=0.18
P
80MHz to 800MHz
d=0.35
P
800MHz to 2.5GHz
Where
P
is the maximum output power rating of the
transmitter in watts (W) according to the transmitter
manufacturer and
d
is the recommended separation
distance in metres (m).
b
Field strengths from fixed RF transmitters, as
determined by an electromagnetic site survey,
a
should be less than the compliance level in each
frequency range .
b
Interference may occur in the vicinity of equipment
marked with the following symbol.
Radiated RF
IEC 61000-4-3
3 V/m
180 MHz - 2.5 GHz
20 V/m
Содержание 35306
Страница 2: ...2 ENGLISH Patient Monitor Operator s Manual ...
Страница 21: ...21 ENGLISH Series III Front view Rear view Left view Right view 11 ...
Страница 22: ...22 ENGLISH Series IV Front view Rear view Left view Right view 11 Extended Module ...
Страница 128: ...128 ENGLISH Oxygenation calculation Calculate result Oxygenation calculation Reference range Unit Result Range ...
Страница 171: ...2 ESPAÑOL Monitor de paciente Manual de operador ...
Страница 189: ...20 ESPAÑOL Serie III Vista delantera Vista posterior Vista lateral izquierda Vista lateral derecha 11 ...
Страница 190: ...21 ESPAÑOL Serie IV Vista delantera Vista posterior Vista lateral izquierda Vista lateral derecha 11 Módulo adicional ...
Страница 343: ...2 FRANÇAIS Moniteur patient Manuel d utilisation ...
Страница 361: ...20 FRANÇAIS Série II Vue de face Vue arrière Connecteur et icône Vue de gauche Vue de droite 13 12 ...
Страница 362: ...21 FRANÇAIS Série III Vue de face Vue arrière Vue de gauche Vue de droite 11 ...
Страница 363: ...22 FRANÇAIS Série IV Vue de face Vue arrière Vue de gauche Vue de droite Panneau avant 11 Module avancé ...
Страница 519: ...2 ITALIANO Monitor Paziente Manuale d Uso ...
Страница 538: ...21 ITALIANO Serie III Vista anteriore Vista posteriore Lato sinistro Lato destro 11 ...
Страница 539: ...22 ITALIANO Serie IV Vista anteriore Vista posteriore Lato sinistro Lato destro 11 Modulo esteso ...