41
ENGLISH
Beat priority:
to set the priority of the source of beat pulse beep and heart beat rate display value, there are 2
options: HR and PR. The factory default is “HR”.
When beat priority is selected to be “HR”
, then the ECG panel displays the current measured heart rate value, and
generates the pulse beep, which are extracted from ECG signal when the heart beat is detected. If the heart rate can not
be obtained, or no heart beat is detected, then the ECG panel shifts to display pulse rate extracted from oximetry signal,
and pulse beep also comes from this source.
When beat priority is selected to be “PR”
, then ECG panel displays the current measured pulse rate value, and generates
the pulse beep, which are extracted from oximetry signal, when the pulse beat is detected. If pulse rate value can not be
obtained and no pulse beat is detected, then the ECG panel shifts to display heart rate extracted from ECG signal, or pulse
beep also comes from this source. Even if the pulse rate value is displayed while the priority of PR is set, if the heart rate
value, which is not displayed, exceeds the preset high/low alarm limits, the alarm will still be activated and alarm event
will be recorded.
The ECG panel will shift to display the high/low alarm limit setting range automatically corresponding to the displayed
heart rate or pulse rate value.
Use the last patient info. When power on
: tick it means that the patient document of the last time will be used as
the Current Patient Document when powered on, i.e. the patient is not changed. Otherwise, system will use the
Default Patient Document (PID:0000000, Noname) as the Current Patient Document when powered on.
Touchscreen Calibration (optional):
to enter into touchscreen calibration window. There are 5 cross cursors “ ” on
the screen, tab the cross point “ ” of the 5 cross cursors with the stylus respectively to finish the calibration.
3.8.2 Setting the Date and Time
Step 1: Select “Menu”
→
“System Settings”
→
“Date/Time”.
Month/Day/Year/Hour/Minute/Second:
to set system date and time.
Date format:
to set date format of the system. There are 4 options: “YYYY-MM-DD”, “YYYY.MM.DD”, “MM/DD/
YYYY”, “DD/MM/YYYY”.
Current time:
to refresh the current system time.
3.8.3 Network Settings
Select “Menu”
→
“System Settings”
→
“Network”.
Server IP Address:
the IP address used to connecting a central monitoring system server (work station) .
Port:
the port number to which the monitor will connect to the work station in the central monitoring system. Its
setting range is from 6001 to 6064. It can also be used to represent the patient bed number connecting to the work
station. For example, the port number means the monitor is assigned to the bed number 2 in the CMS. Our work
station can connect to up to 64 bedside monitors, so please set the port between 6001and 6064. Press “OK” to make
the new setting effective.
HL7 Protocol:
enable or disable the monitor to export data to CIS/HIS by HL7 protocol. Select it means enable HL7
protocol. The default is “enabled” HL7 protocol.
Local IP Address:
the local IP address for the device.
Содержание 35306
Страница 2: ...2 ENGLISH Patient Monitor Operator s Manual ...
Страница 21: ...21 ENGLISH Series III Front view Rear view Left view Right view 11 ...
Страница 22: ...22 ENGLISH Series IV Front view Rear view Left view Right view 11 Extended Module ...
Страница 128: ...128 ENGLISH Oxygenation calculation Calculate result Oxygenation calculation Reference range Unit Result Range ...
Страница 171: ...2 ESPAÑOL Monitor de paciente Manual de operador ...
Страница 189: ...20 ESPAÑOL Serie III Vista delantera Vista posterior Vista lateral izquierda Vista lateral derecha 11 ...
Страница 190: ...21 ESPAÑOL Serie IV Vista delantera Vista posterior Vista lateral izquierda Vista lateral derecha 11 Módulo adicional ...
Страница 343: ...2 FRANÇAIS Moniteur patient Manuel d utilisation ...
Страница 361: ...20 FRANÇAIS Série II Vue de face Vue arrière Connecteur et icône Vue de gauche Vue de droite 13 12 ...
Страница 362: ...21 FRANÇAIS Série III Vue de face Vue arrière Vue de gauche Vue de droite 11 ...
Страница 363: ...22 FRANÇAIS Série IV Vue de face Vue arrière Vue de gauche Vue de droite Panneau avant 11 Module avancé ...
Страница 519: ...2 ITALIANO Monitor Paziente Manuale d Uso ...
Страница 538: ...21 ITALIANO Serie III Vista anteriore Vista posteriore Lato sinistro Lato destro 11 ...
Страница 539: ...22 ITALIANO Serie IV Vista anteriore Vista posteriore Lato sinistro Lato destro 11 Modulo esteso ...