89
ENGLISH
If air bubble appears in pressure tube, please fill the tube with saline solution again. Air bubble may cause
inaccurate reading.
When doing ICP measurement to the patient who is sitting, please keep the pressure transducer and the top of
patient’s ear at the same level, or it may lead to error reading.
When unplugging the cable from the monitor, be sure to hold the head of the connector and pull it out.
Each time when connecting transducer kit or using a new transducer kit, zero calibration to the IBP transducer
must be carried out.
Before using the tube, cable and/or transducer, make sure that all accessories meet the performance requirements
which is not changed by aging or environmental conditions.
12.3 Setting Up the IBP Measurement
If your monitor is configured with Plugin IBP Module, then you need to make sure that the IBP Plug-and-Play box is
securely inserted. If the box is inserted successfully, then the IBP icon appears on the upper right side of the screen.
Refer to the following description.
If your monitor is configured with internal IBP Module, then the IBP icon appears on the upper right side of the screen.
12.3.1 Understanding IBP Icon and IBP Plugin Box (Optional)
The Plugin box (optional) can be configured with 2 IBP module and one C.O. module, or be configured with 2 IBP modules,
as shown in below figure, or be configured with C.O. module, refer to the monitor in your hand.
For some models of the device, the icon for the IBP plugin box will be displayed on the upper right corner of the monitor,
for example, “
” for IBP and C.O. Plugin box, “
” for IBP Plugin box, “
” for C.O. Plugin box.
Appearance for IBP and C.O. Plugin box
The icons “IBP1” and “IBP2” are the connectors to the IBP1 and IBP2. If the Plugin box is inserted into the left slot,
then “IBPL1” and “IBPL2” will be displayed on the screen of the monitor. The connector will be different because of the
different configuration, please refer to the monitor in your hand.
Содержание 35306
Страница 2: ...2 ENGLISH Patient Monitor Operator s Manual ...
Страница 21: ...21 ENGLISH Series III Front view Rear view Left view Right view 11 ...
Страница 22: ...22 ENGLISH Series IV Front view Rear view Left view Right view 11 Extended Module ...
Страница 128: ...128 ENGLISH Oxygenation calculation Calculate result Oxygenation calculation Reference range Unit Result Range ...
Страница 171: ...2 ESPAÑOL Monitor de paciente Manual de operador ...
Страница 189: ...20 ESPAÑOL Serie III Vista delantera Vista posterior Vista lateral izquierda Vista lateral derecha 11 ...
Страница 190: ...21 ESPAÑOL Serie IV Vista delantera Vista posterior Vista lateral izquierda Vista lateral derecha 11 Módulo adicional ...
Страница 343: ...2 FRANÇAIS Moniteur patient Manuel d utilisation ...
Страница 361: ...20 FRANÇAIS Série II Vue de face Vue arrière Connecteur et icône Vue de gauche Vue de droite 13 12 ...
Страница 362: ...21 FRANÇAIS Série III Vue de face Vue arrière Vue de gauche Vue de droite 11 ...
Страница 363: ...22 FRANÇAIS Série IV Vue de face Vue arrière Vue de gauche Vue de droite Panneau avant 11 Module avancé ...
Страница 519: ...2 ITALIANO Monitor Paziente Manuale d Uso ...
Страница 538: ...21 ITALIANO Serie III Vista anteriore Vista posteriore Lato sinistro Lato destro 11 ...
Страница 539: ...22 ITALIANO Serie IV Vista anteriore Vista posteriore Lato sinistro Lato destro 11 Modulo esteso ...