113
ENGLISH
Gain:
the CO
2
waveform gain. 4 options: X1/2, X1, X2 and X4. The default is X1 for adult and pediatric patient, and
X2 for neonate patient.
X1 waveform scale with base gain X1/2 half scale size of the base gain
X2 twice scale size of the base gain X4 four times scale size of the base gain
Speed:
Respiration waveform sweeping speed of, 2 options 6.25mm/s and 12.5 mm/s. The default is 12.5 mm/s.
Apnea(s):
The timeout setting for apnea alarm (in second).
(1). When AG monitoring is on:
It can be set as any number from 20 to 60 seconds, the step is 1 second. Icon “
“ displays on the
upper right corner of Respiration Panel; When the device has not detected any breathing signal for the specified
time, the “Apnea” alarm prompts on, and alarm sound will be activated. If it is set as OFF, the icon “
“
displays on lower left corner of the Respiration Panel. The default is 20s.
(2). When AG monitoring is off:
refer to section
Changing RESP Setting
.
Source:
Respiration signal source. This item is fixed to be “AG” if AG monitoring function is selected. Otherwise, the
source will be obtained from thoracic impedance measurement from ECG module.
Zeroing:
press it to perform zero resetting.
Default:
resume to the factory default value.
CO
2
Unit:
the unit of CO
2
concentration. 3 options: kPa, mmHg and vol%. 1kPa=7.5mmHg. CO
2
(mmHg) = (CO
2
concentration (vol%) ) * (atmospheric pressure (kPa) ) * 7.5.
Barometric:
the detected atmospheric pressure, which can be set at the same time. The setting range is from
1~999(%), the default is 100(%).
O
2
density:
Oxygen concentration setting.
: the switch for displaying waveform of AA1 and AA2. Select it means the AA1 and
AA2 waveform will be displayed on the screen during monitoring.
PARM:
to enter into Anaesthesia Gas alarm setting. See Section
“Understanding the Alarm Setting”.
Содержание 35306
Страница 2: ...2 ENGLISH Patient Monitor Operator s Manual ...
Страница 21: ...21 ENGLISH Series III Front view Rear view Left view Right view 11 ...
Страница 22: ...22 ENGLISH Series IV Front view Rear view Left view Right view 11 Extended Module ...
Страница 128: ...128 ENGLISH Oxygenation calculation Calculate result Oxygenation calculation Reference range Unit Result Range ...
Страница 171: ...2 ESPAÑOL Monitor de paciente Manual de operador ...
Страница 189: ...20 ESPAÑOL Serie III Vista delantera Vista posterior Vista lateral izquierda Vista lateral derecha 11 ...
Страница 190: ...21 ESPAÑOL Serie IV Vista delantera Vista posterior Vista lateral izquierda Vista lateral derecha 11 Módulo adicional ...
Страница 343: ...2 FRANÇAIS Moniteur patient Manuel d utilisation ...
Страница 361: ...20 FRANÇAIS Série II Vue de face Vue arrière Connecteur et icône Vue de gauche Vue de droite 13 12 ...
Страница 362: ...21 FRANÇAIS Série III Vue de face Vue arrière Vue de gauche Vue de droite 11 ...
Страница 363: ...22 FRANÇAIS Série IV Vue de face Vue arrière Vue de gauche Vue de droite Panneau avant 11 Module avancé ...
Страница 519: ...2 ITALIANO Monitor Paziente Manuale d Uso ...
Страница 538: ...21 ITALIANO Serie III Vista anteriore Vista posteriore Lato sinistro Lato destro 11 ...
Страница 539: ...22 ITALIANO Serie IV Vista anteriore Vista posteriore Lato sinistro Lato destro 11 Modulo esteso ...