6.3.6 Changing Alarm Volume
To set the alarm sound volume.
Step 1: Select “Alarm”
→
“Others”.
Step 2: In “Others”, select “Alarm Volume”. Its setting range is “0 ~ 10”, the step is 1. The default is 5. “10” is the maximum volume.
When the alarm sound is switched off, the monitor will give no audible alarm tones even if a new alarm occurs.
Therefore the user should be very careful about whether to switch off the alarm sound or not.
Do not rely exclusively on the audible alarm system for monitoring. Adjustment of alarm volume to a low level
may result in a hazard to the patient. Always keep the patient under close surveillance.
Pausing or switching off alarms may result in a hazard to the patient, please be very careful.
If occurrence of multiple alarm signal at the same time, the monitor will only show the high priority alarm in the form of
audible alarm and alarm lamp. Meanwhile, all alarm signals information including message descriptions and numerical
highlight etc. will be shown respectively and simultaneously on the screen.
It is suggested that the users should not change the alarm volume lower than the factory default setting if close and
constant attention could not be paid to the patient, otherwise the negligence of alarm event might cause irreversible harm
to the patient.
During the alarm silence period, any new alarm event can activate the audible alarm again and the audible alarm function
resumes normal state.
6.4 Understanding the Alarm Setting
Select “Alarm” button on Statusbar, and select parameter (such ECG, SpO
2
, NIBP, TEMP, etc.). You can review and set alarm
limits, alarm switches, high/low alarm limit, and alarm level for each parameter, as shown in below figure.
Alarm setting window (for full configuration)
Содержание 35306
Страница 2: ...2 ENGLISH Patient Monitor Operator s Manual ...
Страница 21: ...21 ENGLISH Series III Front view Rear view Left view Right view 11 ...
Страница 22: ...22 ENGLISH Series IV Front view Rear view Left view Right view 11 Extended Module ...
Страница 128: ...128 ENGLISH Oxygenation calculation Calculate result Oxygenation calculation Reference range Unit Result Range ...
Страница 171: ...2 ESPAÑOL Monitor de paciente Manual de operador ...
Страница 189: ...20 ESPAÑOL Serie III Vista delantera Vista posterior Vista lateral izquierda Vista lateral derecha 11 ...
Страница 190: ...21 ESPAÑOL Serie IV Vista delantera Vista posterior Vista lateral izquierda Vista lateral derecha 11 Módulo adicional ...
Страница 343: ...2 FRANÇAIS Moniteur patient Manuel d utilisation ...
Страница 361: ...20 FRANÇAIS Série II Vue de face Vue arrière Connecteur et icône Vue de gauche Vue de droite 13 12 ...
Страница 362: ...21 FRANÇAIS Série III Vue de face Vue arrière Vue de gauche Vue de droite 11 ...
Страница 363: ...22 FRANÇAIS Série IV Vue de face Vue arrière Vue de gauche Vue de droite Panneau avant 11 Module avancé ...
Страница 519: ...2 ITALIANO Monitor Paziente Manuale d Uso ...
Страница 538: ...21 ITALIANO Serie III Vista anteriore Vista posteriore Lato sinistro Lato destro 11 ...
Страница 539: ...22 ITALIANO Serie IV Vista anteriore Vista posteriore Lato sinistro Lato destro 11 Modulo esteso ...