Auto
: Active closed loop power control: sends the power control bits needed to
control the AT power to the "Manual Expected Power".
●
All Up
: Sends a sequence of power up bits.
●
All Down
: Sends a sequence of power down bits.
●
Hold
: Sends alternating power up and power down bits to keep the output power in
the vicinity of a constant value.
●
Pattern
: Sends the user specific segment bits executed by the command
.
Option R&S CMW-KS890 is required for "Pattern".
Remote command:
CONFigure:EVDO:SIGN<i>:RPControl:PCBits
Step Size
Defines the power control step size, i.e. the nominal change in mean output power per
single power control bit.
Option R&S CMW-KS890 is required.
Remote command:
CONFigure:EVDO:SIGN<i>:RPControl:SSIZe
Repetition
Specifies whether the power control bit pattern is executed once (single-shot) or repea-
ted continuously and stopped by the
Remote command:
CONFigure:EVDO:SIGN<i>:RPControl:REPetition
Segments
Sets the user specific power control bits. Combine the bits "All Up", "All Down" and
"Alternating" within the four segments of the variable length. These parameters are rel-
evant only for the
Option R&S CMW-KS890 is required.
Remote command:
CONFigure:EVDO:SIGN<i>:RPControl:SEGMent<no>:BITS
CONFigure:EVDO:SIGN<i>:RPControl:SEGMent<no>:LENGth
Run Sequence
Starts and in continuous mode also stops the execution of the user specific pattern set
by
This parameter is relevant only for the
= "Pattern".
Option R&S CMW-KS890 is required.
Remote command:
CONFigure:EVDO:SIGN<i>:RPControl:RUN
2.3.7.8
Time Settings
Defines the CDMA2000 system time settings. For more information refer to 3GPP2
C.S0002.