![Rohde & Schwarz CMW-KE100 User Manual Download Page 362](http://html.mh-extra.com/html/rohde-and-schwarz/cmw-ke100/cmw-ke100_user-manual_1477807362.webp)
1xEV-DO Measurements
R&S
®
CMW-KM8xx/-KS8xx
362
User Manual 1173.9511.02 ─ 12
READ:EVDO:MEAS<i>:MEValuation:TRACe:ACP:EXTended:MAXimum[:RELative]?
FETCh:EVDO:MEAS<i>:MEValuation:ACP:CURRent?
...................................................... 365
FETCh:EVDO:MEAS<i>:MEValuation:ACP:AVERage?
......................................................365
FETCh:EVDO:MEAS<i>:MEValuation:ACP:MAXimum?
.....................................................365
READ:EVDO:MEAS<i>:MEValuation:ACP:CURRent?
....................................................... 365
READ:EVDO:MEAS<i>:MEValuation:ACP:AVERage?
.......................................................365
READ:EVDO:MEAS<i>:MEValuation:ACP:MAXimum?
...................................................... 365
CALCulate:EVDO:MEAS<i>:MEValuation:ACP:CURRent?
.................................................365
CALCulate:EVDO:MEAS<i>:MEValuation:ACP:AVERage?
................................................ 365
CALCulate:EVDO:MEAS<i>:MEValuation:ACP:MAXimum?
............................................... 365
CALCulate:EVDO:MEAS<i>:MEValuation:TRACe:ACP:CURRent:ABSolute?
CALCulate:EVDO:MEAS<i>:MEValuation:TRACe:ACP:AVERage:ABSolute?
CALCulate:EVDO:MEAS<i>:MEValuation:TRACe:ACP:MAXimum:ABSolute?
FETCh:EVDO:MEAS<i>:MEValuation:TRACe:ACP:CURRent:ABSolute?
FETCh:EVDO:MEAS<i>:MEValuation:TRACe:ACP:AVERage:ABSolute?
FETCh:EVDO:MEAS<i>:MEValuation:TRACe:ACP:MAXimum:ABSolute?
READ:EVDO:MEAS<i>:MEValuation:TRACe:ACP:CURRent:ABSolute?
READ:EVDO:MEAS<i>:MEValuation:TRACe:ACP:AVERage:ABSolute?
READ:EVDO:MEAS<i>:MEValuation:TRACe:ACP:MAXimum:ABSolute?
Returns the adjacent channel absolute power measured in dBm at a series of frequen-
cies.
The frequencies are determined by the offset values defined via the commands
CONFigure:EVDO:MEAS<i>:MEValuation:ACP:FOFFsets:LOWer
and
CONFigure:EVDO:MEAS<i>:MEValuation:ACP:FOFFsets:UPPer
. All defined
offset values are considered (irrespective of their activation status).
The current, average and maximum traces can be retrieved.
Return values:
<1_Reliability>
<2_PowOffsetM9> ...
<11_PowOffsetM0>
<12_PowCarrier>
<13_PowOffsetP0> ...
<22_PowOffsetP9>
The power offsets <PowOffsetPn> (<PowOffsetMn>) refer to the
average absolute power at the carrier frequency plus (minus) the
frequency offset value number n.
Range:
-100 dBm to 20 dBm
Default unit: dBm
Example:
See
Multi Evaluation: Performing Single-Shot Measurements
Usage:
Query only
Firmware/Software:
V3.2.70
For additional information concerning syntax elements and returned values refer to
Conventions and General Information
.
FETCh:EVDO:MEAS<i>:MEValuation:TRACe:ACP:CURRent[:RELative]?
FETCh:EVDO:MEAS<i>:MEValuation:TRACe:ACP:AVERage[:RELative]?
FETCh:EVDO:MEAS<i>:MEValuation:TRACe:ACP:MAXimum[:RELative]?
READ:EVDO:MEAS<i>:MEValuation:TRACe:ACP:CURRent[:RELative]?
Command Reference