1xEV-DO Measurements
R&S
®
CMW-KM8xx/-KS8xx
261
User Manual 1173.9511.02 ─ 12
3.2.9.3
OLTR: Results
The OLTR view shows the DUT's responses to the power down and power up steps of
a single OLTR measurement sequence. The responses are recorded as power vs.
time traces, with power values (in dB) relative to the corresponding reference power
level. The direction of the first power step is displayed above, the reference power of
both steps is displayed below the diagrams. The red lines represent the response corri-
dor specified as minimum standard in 3GPP2 C.S0011.
Limit violations in particular time interval are indicated underneath the time axis. The
limit mask used for this measurement represent the minimum standard for "Time
Response of Open Loop Power Control" specified in 3GPP2 C.S0033.
Fig. 3-19: 1xEV-DO OLTR measurement: one measurement sequence
Each step trace consists of 60 equidistant values for the reference power interval, 640
equidistant values for the OLTR evaluation interval and 20 equidistant indicators of the
OLTR limit violation.
3.2.10 Selecting and Modifying Views
Use the "Display" parameters to select the views and to change the appearance and
contents of the views. Depending on the selected view the following "Display" hotkeys
are available at the bottom of the GUI: