1xEV-DO Signaling
R&S
®
CMW-KM8xx/-KS8xx
160
User Manual 1173.9511.02 ─ 12
Manual operation:
See
CONFigure:EVDO:SIGN<i>:RFPower:EXPected?
Queries the calculated value of the expected input power from the AT.
The input power range is stated in the data sheet.
Return values:
<ExpNomPower>
*RST:
0 dBm
Default unit: dBm
Example:
See
chapter 2.4.1.1, "Specifying General Settings"
Usage:
Query only
Firmware/Software:
V1.0.15.0
Manual operation:
See
CONFigure:EVDO:SIGN<i>:RFPower:MODE:AWGN
<AWGNMode>
Selects the operating mode of the AWGN generator. The AWGN level range
(
CONFigure:EVDO:SIGN<i>:RFPower:LEVel:AWGN
) depends on the operating
mode.
Parameters:
<AWGNMode>
NORMal | HPOWer
Normal or High Power AWGN mode
*RST:
NORM
Example:
See
Firmware/Software:
V1.0.15.0
Options:
High Power mode requires R&S CMW-KS890
Manual operation:
See
CONFigure:EVDO:SIGN<i>:RFPower:LEVel:AWGN
<AWGNLevel>
Sets/gets the state and level of the AWGN generator relative to the "1xEV-DO Power"
(
CONFigure:EVDO:SIGN<i>:RFPower:EVDO
). The AWGN level range depends on
the operating mode of the AWGN generator (
).
With the "set" command the AWGN generator can be turned OFF or ON and the level
can be set. If the level is set, the generator is automatically turned ON.
The query returns either the OFF state or the level if the generator state is ON.
Command Reference