1xEV-DO Signaling
R&S
®
CMW-KM8xx/-KS8xx
104
User Manual 1173.9511.02 ─ 12
Single-shot is preferable if only a single measurement result is required under fixed
conditions, which is typical for remote-controlled measurements. Continuous mode is
suitable for monitoring the evolution of the measurement results in time and observe
how they depend on the measurement configuration, which is typically done in manual
control. The reset/preset values therefore differ from each other.
For PER and Throughput measurements, independent repetition types can be
selected.
Remote command:
CONFigure:EVDO:SIGN<i>:RXQuality:PER:REPetition
CONFigure:EVDO:SIGN<i>:RXQuality:THRoughput:REPetition
Forward/Reverse Link PER > Stop Condition
Specifies the conditions for an early termination of the measurement:
●
None
: The measurement is performed according to its "Repetition" mode and
"Transport Blocks", irrespective of the limit check results.
●
On Limit Failure
: The measurement is stopped as soon as one of the limits is
exceeded, irrespective of the repetition mode set. If no limit failure occurs, it is per-
formed according to its "Repetition" mode and "Transport Blocks" settings. Use this
value for measurements that are essentially intended for checking limits, e.g. pro-
duction tests.
Remote command:
CONFigure:EVDO:SIGN<i>:RXQuality:FLPer:SCONdition
CONFigure:EVDO:SIGN<i>:RXQuality:RLPer:SCONdition
Forward/Reverse Link PER > Max Test Packets Sent
Defines the number of test packets to be sent by the R&S
CMW/AT for the respective
measurement (statistics cycle, single-shot measurement).
See also: "Statistical Results" in the R&S
CMW user manual, chapter "System Over-
view"
Remote command:
CONFigure:EVDO:SIGN<i>:RXQuality:FLPer:MTPSent
CONFigure:EVDO:SIGN<i>:RXQuality:RLPer:MPSent
Forward/Reverse Link Throughput > Max Frames
Defines the maximum duration of the "Forward Link Throughput" measurement as a
number of 26.66 ms CDMA frames (statistics cycle, single-shot measurement).
See also: "Statistical Results" in the R&S
CMW user manual, chapter "System Over-
view"
Remote command:
Limits (for PER)
In the "Limits" section you can configure the PER limits including the limits for confi-
dence testing, see
chapter 2.2.14.4, "Statistical PER Tests (Confidence Testing)"
GUI Reference