1xEV-DO Signaling
R&S
®
CMW-KM8xx/-KS8xx
123
User Manual 1173.9511.02 ─ 12
CONFigure:EVDO:SIGN:RXQuality:LIMit:FLPer:CLEVel 95
// *************************************************************************
// Reverse Link PER measurement configuration: lLet the AT send at most
// 1000 test application MAC packets. Stop if either these 1000 packets
// have been sent or the configured confidence level has been reached.
// Set the maximum tolerable PER to 3% and the minimum required confidence
// level to 95%
// *************************************************************************
CONFigure:EVDO:SIGN:RXQuality:RLPer:MPSent 1000
CONFigure:EVDO:SIGN:RXQuality:RLPer:SCONdition MCLexceeded
CONFigure:EVDO:SIGN:RXQuality:LIMit:RLPer:MPER 3
CONFigure:EVDO:SIGN:RXQuality:LIMit:FLPer:CLEVel 95
2.4.2.3
Configure Performance (Throughput) Measurements
// *************************************************************************
// Define the repetition mode for both directions.
// *************************************************************************
CONFigure:EVDO:SIGN:RXQuality:THRoughput:REPetition CONTinuous
// *************************************************************************
// Set the duration
// *************************************************************************
CONFigure:EVDO:SIGN:RXQuality:FLPFormance:MFRames 1000
CONFigure:EVDO:SIGN:RXQuality:RLPFormance:MFRames 1000
2.4.2.4
Setting up the Connection
To set up the connection see
chapter 2.4.1.8, "Setting up a Connection"
2.4.2.5
Perform Forward Link Measurements
// *************************************************************************
// Make sure the "Forward" test application is running, enable forward
// link PER measurement, start the PER measurement (in single-shot mode),
// wait until sufficient data has been collected, retrieve scalar
// results for the available carriers.
// Query the measurement state (should be "OK").
// Query the measurement state (should be "RDY").
// *************************************************************************
CONFigure:EVDO:SIGN:RXQuality:RESult:FLPer ON
INIT:EVDO:SIGN:PER
CONFigure:EVDO:SIGN:RXQuality:CARRier:SELect 0
FETCh:EVDO:SIGN:RXQuality:FLPer?
CONFigure:EVDO:SIGN:RXQuality:CARRier:SELect 1
FETCh:EVDO:SIGN:RXQuality:FLPer?
Programming