![Rohde & Schwarz CMW-KE100 User Manual Download Page 808](http://html.mh-extra.com/html/rohde-and-schwarz/cmw-ke100/cmw-ke100_user-manual_1477807808.webp)
CDMA2000 Measurements
R&S
®
CMW-KM8xx/-KS8xx
808
User Manual 1173.9511.02 ─ 12
<RConfig>
R12Q | R36H | R3Q
Selects the radio configuration which determines for example
the modulation type.
Note
that if
RetriggerOption=OFF
for all segments (i.e. in
"trigger once" mode), the radio configuration is inherited from the
multi evaluation measurement (see
RConfig
values specified for
the individual segments.
R12Q:
RC1 or 2 (O-QPSK)
R36H:
RC3 to 6 (H-PSK)
R3Q:
RC3 (QPSK)
*RST:
R36H
<CMWSConnector> Optional parameter, as alternative to the command
CONFigure:CDMA:MEAS<i>:MEValuation:LIST:
SEGMent<no>:CMWS:CONNector
*RST:
R11
Example:
See
Specifying Global Measurement Settings
Firmware/Software:
V1.0.15.20
V1.0.15.23: new parameters <RetriggerOption> and <RConfig>
V3.2.70: added <CMWSConnector>
Options:
R&S CMW-KM012
CONFigure:CDMA:MEAS<i>:MEValuation:LIST:SEGMent<no>:MODulation
<StatisticLength>, <EnableEVM>, <EnableMagError>, <EnablePhaseErr>,
<EnableWaveQual>, <EnableIQError>, <EnableChPow>, <EnableChTimeOff>,
<EnableChPhseOff>, <EnableWBNBPow>, <EnableFreqErr>
Defines the statistical length for
AVERage
,
MAXimum
,
MINimum
and
SDEViation
cal-
culation and enables the calculation of the different modulation results in segment
<no>; see
.
Suffix:
<no>
.
1..200
The segment number must not exceed the total number of seg-
ments measured.
Parameters:
<StatisticLength>
The statistical length is limited by the segment length (set with
CONFigure:CDMA:MEAS<i>:MEValuation:LIST:
SEGMent<no>:SETup
) and depends on the used trigger:
statistical length
≤
segment length - 2
for "IF Auto Sync" trigger,
statistical length
≤
segment length - 1
for other triggers
*RST:
1
Command Reference
深圳德标仪器
135-1095-0799