![Rohde & Schwarz CMW-KE100 User Manual Download Page 811](http://html.mh-extra.com/html/rohde-and-schwarz/cmw-ke100/cmw-ke100_user-manual_1477807811.webp)
CDMA2000 Measurements
R&S
®
CMW-KM8xx/-KS8xx
811
User Manual 1173.9511.02 ─ 12
Parameters:
<RetriggerMode>
ONCE | SEGMent
ONCE:
Trigger only once. Every segment is measured irrespec-
tive of the setting of the parameter
<RetriggerOption>
from
the segment (
CONFigure:CDMA:MEAS<i>:MEValuation:
The trigger is rearmed only after the measurement is stopped
and re-started.
SEGM:
The measurement starts after the first trigger event and
continues as long as no segment is reached that requires a ret-
rigger (
CONFigure:CDMA:MEAS<i>:MEValuation:LIST:
) . This mode is recommended for statis-
tic counts where retriggering can compensate a possible time
drift of the MS.
*RST:
ONCE
Example:
See
Firmware/Software:
V1.0.15.23
Options:
R&S CMW-KM012
5.5.4.5
Limits (Power and Modulation)
The following commands define limits for results which characterize power require-
ments and modulation accuracy.
CONFigure:CDMA:MEAS<i>:MEValuation:LIMit:POWer
.................................................... 811
CONFigure:CDMA:MEAS<i>:MEValuation:LIMit:EVMagnitude
........................................... 812
CONFigure:CDMA:MEAS<i>:MEValuation:LIMit:MERRor
.................................................. 812
CONFigure:CDMA:MEAS<i>:MEValuation:LIMit:PERRor
...................................................813
CONFigure:CDMA:MEAS<i>:MEValuation:LIMit:IQOFfset
..................................................813
CONFigure:CDMA:MEAS<i>:MEValuation:LIMit:IQIMbalance
............................................ 813
CONFigure:CDMA:MEAS<i>:MEValuation:LIMit:CFERror
.................................................. 814
CONFigure:CDMA:MEAS<i>:MEValuation:LIMit:TTERror
.................................................. 814
CONFigure:CDMA:MEAS<i>:MEValuation:LIMit:WFQuality
................................................814
CONFigure:CDMA:MEAS<i>:MEValuation:LIMit:POWer
<MinPower>, <MaxPower>
Defines limits for the mobile station (MS) power.
Parameters:
<MinPower>
Range:
-128 dBm to 50 dBm
*RST:
-40 dBm
Default unit: dBm
Additional parameters: OFF | ON (disables the limit check | ena-
bles the limit check using the previous/default limit values)
Command Reference
135-1095-0799