1xEV-DO Signaling
R&S
®
CMW-KM8xx/-KS8xx
166
User Manual 1173.9511.02 ─ 12
2.5.6.7
Reverse Power Control
The following commands provide closed loop power control settings.
CONFigure:EVDO:SIGN<i>:RPControl:PCBits
..................................................................166
CONFigure:EVDO:SIGN<i>:RPControl:SSIZe
...................................................................166
CONFigure:EVDO:SIGN<i>:RPControl:REPetition
............................................................ 167
CONFigure:EVDO:SIGN<i>:RPControl:SEGMent<no>:BITS
.............................................. 167
CONFigure:EVDO:SIGN<i>:RPControl:SEGMent<no>:LENGth
..........................................167
CONFigure:EVDO:SIGN<i>:RPControl:RUN
.....................................................................168
CONFigure:EVDO:SIGN<i>:RPControl:PCBits
<PCBits>
Defines a power control bit pattern which the R&S
CMW transmits in order to control
the transmitter output power of the AT.
Parameters:
<PCBits>
AUTO | AUP | ADOWn | HOLD | PATTern
AUTO:
Active closed loop power control
AUP:
Power up bits
ADOW:
Power down bits
HOLD:
Alternating power up and power down bits
PATT
: Sends the user specific segment bits executed by
CONFigure:EVDO:SIGN<i>:RPControl:RUN
*RST:
AUTO
Example:
See
Configuring Reverse Power Control Setup
Firmware/Software:
V1.0.15.0
V3.2.80: added "PATT"
Options:
R&S CMW-K890 required for PATTern
Manual operation:
See
CONFigure:EVDO:SIGN<i>:RPControl:SSIZe
<SSize>
Gets/sets the power control step size, i.e. the nominal change in mean output power
per single power control bit.
Parameters:
<SSize>
Range:
0.5 dB to 1 dB
*RST:
1 dB
Default unit: dB
Example:
See
Configuring Reverse Power Control Setup
Firmware/Software:
V1.0.15.0
Options:
R&S CMW-KS890
Manual operation:
See
Command Reference