1xEV-DO Signaling
R&S
®
CMW-KM8xx/-KS8xx
151
User Manual 1173.9511.02 ─ 12
Parameters:
<Seed>
Range:
0 to 9
*RST:
0
Example:
See
Firmware/Software:
V3.2.80
Options:
R&S CMW-KE100 and R&S CMW-KE800
Manual operation:
See
CONFigure:EVDO:SIGN<i>:FADing:FSIMulator:ILOSs:LOSS
<InsertionLoss>
Sets the insertion loss for the fading simulator.
A setting is only allowed in
USER
CONFigure:EVDO:SIGN<i>:FADing:
).
Parameters:
<InsertionLoss>
Range:
0 dB to 18 dB
*RST:
0 dB
Default unit: dB
Example:
See
Firmware/Software:
V3.2.80
Options:
R&S CMW-KE100 and R&S CMW-KE800
Manual operation:
See
CONFigure:EVDO:SIGN<i>:FADing:FSIMulator:ILOSs:MODE
<InsertLossMode>
Sets the insertion loss mode.
Parameters:
<InsertLossMode>
NORMal | USER
NORMal
: the insertion loss is determined by the fading profile
USER
: the insertion loss can be adjusted by the user
*RST:
NORM
Example:
See
Firmware/Software:
V3.2.80
Options:
R&S CMW-KE100 and R&S CMW-KE800
Manual operation:
See
CONFigure:EVDO:SIGN<i>:FADing:FSIMulator:RESTart
Restarts the fading process in
MANual
mode (see
FADing:FSIMulator:RESTart:MODE
).
Usage:
Event
Command Reference