![Rohde & Schwarz CMW-KE100 User Manual Download Page 212](http://html.mh-extra.com/html/rohde-and-schwarz/cmw-ke100/cmw-ke100_user-manual_1477807212.webp)
1xEV-DO Signaling
R&S
®
CMW-KM8xx/-KS8xx
212
User Manual 1173.9511.02 ─ 12
Example:
See
Perform Forward Link Measurements
Usage:
Query only
Firmware/Software:
V2.1.20
Options:
R&S CMW-KS890 required for Throughput measurements
Manual operation:
See
FETCh:EVDO:SIGN<i>:PER:STATe:ALL?
FETCh:EVDO:SIGN<i>:THRoughput:STATe:ALL?
Queries the main measurement state and the measurement substates. Both measure-
ment substates are relevant for running measurements only. Use
FETCh:...:STATe?
to query the main measurement state only. Use
INITiate...,
STOP..., ABORt...
to change the measurement state.
See also: "Measurement Control" in the R&S
CMW user manual, chapter "Remote
Control"
Return values:
<MainState>
OFF | RDY | RUN
OFF:
measurement switched off, no resources allocated, no
results available (when entered after
STOP...
)
RDY:
measurement has been terminated, valid results may be
available
RUN:
measurement running (after
INITiate..., READ...
),
synchronization pending or adjusted, resources active or queued
<SyncState>
PEND | ADJ | INV
PEND:
waiting for resource allocation, adjustment, hardware
switching ("pending")
ADJ:
all necessary adjustments finished, measurement running
("adjusted")
INV:
not applicable because <main_state>: OFF or RDY ("inva-
lid")
<RessourceState>
QUE | ACT | INV
QUE:
measurement without resources, no results available
("queued")
ACT:
resources allocated, acquisition of results in progress but
not complete ("active")
INV:
not applicable because <main_state>: OFF or RDY ("inva-
lid")
Usage:
Query only
Firmware/Software:
V2.1.20
Options:
R&S CMW-KS890 required for Throughput measurements
Manual operation:
See
Command Reference