![Rohde & Schwarz CMW-KE100 User Manual Download Page 295](http://html.mh-extra.com/html/rohde-and-schwarz/cmw-ke100/cmw-ke100_user-manual_1477807295.webp)
1xEV-DO Measurements
R&S
®
CMW-KM8xx/-KS8xx
295
User Manual 1173.9511.02 ─ 12
// single-shot measurement, power step 30 dB, direction up,
// reference power interval 20 ms, guard interval 50 ms,
// measure on exception and initial lower OLTR limit.
// *************************************************************************
CONFigure:EVDO:MEAS:OLTR:TOUT 5
CONFigure:EVDO:MEAS:OLTR:SEQuence 5
CONFigure:EVDO:MEAS:OLTR:PSTep 30
CONFigure:EVDO:MEAS:OLTR:PSTep:DIRection UP
CONFigure:EVDO:MEAS:OLTR:RPINterval 12
CONFigure:EVDO:MEAS:OLTR:RPINterval:TIME?
CONFigure:EVDO:MEAS:OLTR:GINTerval 30
CONFigure:EVDO:MEAS:OLTR:GINTerval:TIME?
CONFigure:EVDO:MEAS:OLTR:MOEXception ON
CONFigure:EVDO:MEAS:OLTR:LIMit:ILOWer -2
3.4.1.8
OLTR: Single-Shot Measurement
// *************************************************************************
// Start single-shot measurement for the open loop time response for
// BS power up (resulting in DUT's power down), and return the result
// trace and the limit violation evaluation for the second sequence
// (of the five configured), and also the trace of the reference
// power interval. Query the measurement state (should be "RDY").
// Query measurement state and substates (should be "RUN,ADJ,ACT").
// *************************************************************************
INITiate:EVDO:MEAS:OLTR
FETCh:EVDO:MEAS:OLTR:SEQuence2:TRACe:DOWN?
FETCh:EVDO:MEAS:OLTR:SEQuence2:TRACe:DOWN:STATe?
FETCh:EVDO:MEAS:RPINterval:SEQuence2:TRACe:DOWN?
FETCh:EVDO:MEAS:OLTR:STATe?
FETCh:EVDO:MEAS:OLTR:STATe:ALL?
3.4.2 Using 1xEV-DO List Mode
The 1xEV-DO multi evaluation measurement is programmed as follows:
●
The measurement is controlled by SCPI commands with the following syn-
tax:
...EVDO:MEAS:MEValuation:LIST...
●
Use general commands of the type
...:EVDO:MEAS:...
(no
:MEValuation
/
no
:OLTR
mnemonic) to define the signal routing and perform RF and analyzer
settings.
●
After a
*RST
, the measurement is switched off and list mode is disabled. Use
CONFigure:EVDO:MEAS:MEValuation:LIST ON
to enable the list mode and
INIT:EVDO:MEAS:MEValuation
to initiate a single-shot measurement.
●
Use
FETCh:EVDO:MEAS:MEValuation:LIST:...?
commands to retrieve the
results.
Programming
深圳德标仪器