1xEV-DO Signaling
R&S
®
CMW-KM8xx/-KS8xx
27
User Manual 1173.9511.02 ─ 12
For details to connection setup refer to
chapter 2.2.2, "Initiating Signaling Tests"
2.2.8.2
Initiating the Handoff
After you have completed the connection setup, perform the following steps to initiate a
handoff.
1. Press the "Handoff ..." hotkey in the 1xEV-DO signaling application.
The handoff configuration dialog opens.
2. The 1xEV-DO signaling application is pre-selected as handoff target. Configure the
destination band and the channels of the carriers 0 to 2 and enable pilot channels if
needed.
Make sure the active carriers don't overlap, i.e. their center frequencies have to be
separated by at least CDMA system bandwidth (1.23 MHz for band class 3 JTACS,
1.25 MHz for all other band classes).
General Description