![Rohde & Schwarz CMW-KE100 User Manual Download Page 72](http://html.mh-extra.com/html/rohde-and-schwarz/cmw-ke100/cmw-ke100_user-manual_1477807072.webp)
1xEV-DO Signaling
R&S
®
CMW-KM8xx/-KS8xx
72
User Manual 1173.9511.02 ─ 12
Application
Properties
To be used for...
Forward and Reverse Test
Test data transmission on both
the forward and reverse link. No
additional information can be
transmitted.
Stress tests for simultaneous AT
receiver and transmitter operation
at variable data rates
Packet
Data transfers between DUT and
DAU (Data Application Unit)
User defined data transfers and
throughput measurements for
"real-life" traffic
The packet application can only be de-/selected in signal OFF state.
The protocol parameters appear in the configuration dialog.
To make measurements using the test application protocol, it is required to enable the
test application on a particular stream (1-3).
Remote command:
CONFigure:EVDO:SIGN<i>:APPLication:MODE
CONFigure:EVDO:SIGN<i>:APPLication
Preferred Packet Mode
Sets the preferred standard to be signaled to the AT during the connection setup.
Remote command:
CONFigure:EVDO:SIGN<i>:LAYer:APPLication:PACKet:MODE
2.3.8.2
Application Layer - Release 0
The following parameters configure the test application for network release 0 (physical
layer subtype 0/1) signals.
For background information refer to
chapter 2.2.10, "Test Application Protocols"
Fig. 2-33: Application layer settings for network release 0
............................................................................................ 73
.........................................................................................74
RTAP Configuration > Min / Max Data Rate
................................................................. 74
GUI Reference