CDMA2000 Signaling
R&S
®
CMW-KM8xx/-KS8xx
593
User Manual 1173.9511.02 ─ 12
The
<instance>
is relevant for instruments supporting several instances of the same
firmware application. It can be omitted if the instrument supports only one instance, or
to address the first instance.
See also: "Firmware Applications" in the R&S
CMW user manual, chapter "Remote
Control"
4.5.1.2
FETCh, READ and CALCulate Commands
All commands are used to retrieve measurement results:
●
FETCh...
returns the results of the current measurement cycle (single-shot mea-
surement) after they are valid.
FETCh...
must be used after the measurement
has been started (
INITiate...
, measurement states RUN or RDY).
●
READ...
starts a new single-shot measurement and returns the results.
●
CALCulate...
returns one limit check result per
FETCh
result:
–
OK
: The
FETCh
result is located within the limits or no limit has been defined/
enabled for this result.
–
ULEU
(User limit exceeded upper): An upper limit is violated. The
FETCh
result
is located above the limit.
–
ULEL
(User limit exceeded lower): A lower limit is violated. The
FETCh
result is
located below the limit.
See also: "Retrieving Measurement Results" in the R&S
CMW user manual, chapter
"Remote Control"
4.5.1.3
Values for Signal Path Selection
To select an RF path, you must specify an RF connector and a TX or RX module (con-
verter). To select an I/Q path, you specify a DIG IQ connector.
Which connectors and modules can be specified in a command, depends on the instal-
led hardware and the active sub-instrument or instance <i>.
This section lists all values available for path selection. Depending on your configura-
tion, only a subset may be relevant for you. Virtual connector names are only relevant
for setting commands. Queries return the physical connector names.
Additional information is available in the base software documentation. It describes typ-
ical instrument configurations with the allowed RF connector - TX/RX module combina-
tions and the mapping of virtual connector names to physical connectors.
See also: "Signal Path Settings" in the R&S
CMW user manual, chapter "Remote Con-
trol"
RF path selection values:
●
RX module:
RX1 | RX2 | RX3 | RX4
●
TX module:
TX1 | TX2 | TX3 | TX4
●
RX connector:
Command Reference