1xEV-DO Signaling
R&S
®
CMW-KM8xx/-KS8xx
56
User Manual 1173.9511.02 ─ 12
Crest Factor (Out)
Indicates the crest factor of the baseband signal, i.e. the ratio of peak to average base-
band power. The average power is calculated for time intervals with active forward link
traffic channel timeslots only.
Use the displayed crest factor value to configure the baseband input of the connected
instrument.
Remote command:
SENSe:EVDO:SIGN<i>:IQOut:PATH<n>?
Baseband Level (In)
Indicates the nominal RMS level of the baseband signal during a call (connection
established).
Configure the baseband level so that it matches the baseband output of the connected
instrument.
Remote command:
CONFigure:EVDO:SIGN<i>:IQIN:PATH<n>
2.3.6 Internal Fading
This branch of the configuration tree is only visible if a fading scenario is selected
("Standard Cell Fading") and the fading source is set to "Internal".
For general prerequisites, required options and background information see
Fig. 2-18: Internal fading settings
............................................................................................. 58
............................................................................................ 59
2.3.6.1
Fading Simulator
The following parameters allow to enable and set up the fading simulator. For back-
ground information see
chapter 2.2.6.1, "Fading Simulator"
GUI Reference