1xEV-DO Measurements
R&S
®
CMW-KM8xx/-KS8xx
306
User Manual 1173.9511.02 ─ 12
ROUTe:EVDO:MEAS<i>:SCENario?
Returns the active scenario.
Return values:
<Scenario>
SALone | CSPath
SALone
: Standalone (Non Signaling)
CSPath
: Combined Signal Path
*RST:
SAL
Usage:
Query only
Firmware/Software:
V1.0.15.20
Manual operation:
See
ROUTe:EVDO:MEAS<i>:SCENario:SALone
<RXConnector>, <RXConverter>
Activates the standalone scenario and selects the RF input path for the measured RF
signal.
For possible connector and converter values, see
chapter 3.5.1.4, "Values for RF Path
Parameters:
<RXConnector>
RF connector for the input path
<RXConverter>
RX module for the input path
Example:
See
Specifying General Measurement Settings
Firmware/Software:
V1.0.15.20
Manual operation:
See
ROUTe:EVDO:MEAS<i>:SCENario:CSPath
<Master>
Activates the combined signal path scenario and selects a master. The master controls
the signal routing, analyzer settings and AT signal info settings while the combined sig-
nal path scenario is active. Configure the connector and converter settings via
ROUTe:EVDO:SIGN<i>:SCENario:...
.
Depending on the installed options the set of masters available at your instrument may
differ from the values listed below. A complete list of all supported values can be dis-
played using
ROUTe:EVDO:MEAS<i>:SCENario:CATalog:CSPath?
Parameters:
<Master>
String parameter containing the master application, e.g. '1xEV-
DO
Sig1' or '1xEV-DO
Sig2'
Firmware/Software:
V1.0.15.20
Manual operation:
See
Command Reference