![Rohde & Schwarz CMW-KE100 User Manual Download Page 13](http://html.mh-extra.com/html/rohde-and-schwarz/cmw-ke100/cmw-ke100_user-manual_1477807013.webp)
1xEV-DO Signaling
R&S
®
CMW-KM8xx/-KS8xx
13
User Manual 1173.9511.02 ─ 12
instance 1 in parallel, but not instance 1 of both applications. Exception is hybrid mode,
for details refer to
Test Setup for Standard Cell Scenario
....................................................................13
Test Setup for Hybrid Mode Scenario
.....................................................................13
Test Setup for External Fading Scenarios
.............................................................. 14
Test Setup for Internal Fading Scenarios
................................................................15
2.2.1.1
Test Setup for Standard Cell Scenario
The AT under test can be connected to one of the bidirectional RF connectors (e.g. RF
1 COM or RF 2 COM) at the front panel of the R&S
CMW. No additional cabling and
no external trigger is needed. The input level ranges of all RF COM connectors are
identical.
The bidirectional RF connection between the tester and the DUT carries both the for-
ward and the reverse signal:
●
The R&S
CMW transmits the forward (access network) signal to which the access
terminal can synchronize in order to perform session negotiation and establish the
connection. The forward signal is used to transfer signaling messages and user
data to the AT.
●
The AT transmits a reverse signal that the R&S
CMW can receive and decode in
order to set up a connection and perform various measurements.
See also: "RF Connectors" in the R&S
CMW user manual, chapter "Getting Started"
2.2.1.2
Test Setup for Hybrid Mode Scenario
A test setup for a connection with CDMA2000 1xRTT and CDMA2000 1xEV-DO
involves two forward and two reverse signals. The two signals must be transmitted via
different RF converters.
For the HW specification installed at at the R&S
Use a bidirectional RF connector for the first forward and reverse signal and a separate
RF connector for the second forward and reverse link. Connect both RF connectors to
an external combiner and connect the combiner to the DUT according to the figure
below.
General Description
深圳德标仪器
135-1095-0799